34941A And 34942A Scpi Programming Examples - Agilent Technologies Switch/Measure User Manual

Multifunction switch/measure unit
Table of Contents

Advertisement

7
RF Multiplexer Switch Modules

34941A and 34942A SCPI Programming Examples

The programming examples below provide you with SCPI command
examples to use for actions specific to the RF MUX switch modules.
The slot and channel addressing scheme used in these examples follow
the form sccc where s is the mainframe slot number (1 through 8) and
ccc is the channel number. For information on specific configurations,
refer to the simplified schematic on
For complete information on the SCPI commands used to program the
34980A, refer to the Agilent 34980A Programmer's Reference contained
on the 34980A Product Reference CD. For example programs, also refer
to the 34980A Product Reference CD.
Example: Closing channels
modules. You cannot open channels. When you close a channel, any
already- closed channels automatically open. With this "one- step"
operation, the relays switch in the proper order that avoids momentary
connection of the wrong input to the multiplexer output. The following
command closes channel 03 on Bank 1 of an RF MUX module installed
in slot 5.
ROUTe:CLOSe (@5103)
Example: Querying channels for open or close state The following commands
returns the close or open state of channel 33 of a module installed in
slot 5.
ROUT:CLOSe? (@5033)
ROUT:OPEN? (@5033)
Example: Querying the system for module identify The following command
returns the identify of the module installed in slot 7.
SYSTem:CTYPe? 7
Example: Reading the cycle count for a relay On these modules, the signal
path to the COM terminal consists of two relays. See the simplified
schematic on
for both relays, and returns the greater of the two. The cycle count for
the two channels within the same physical relay package (as indicated by
the dashed lines in the simplified schematic) will always be equal.
Therefore, the cycle count for Channels 101 and 102 will always be equal.
The following statement reads back the number of completed cycles for
the channels 101 and 202 on a module installed in slot 6.
DIAGnostic:RELay:CYCLes? (@6101,6202)
178
You can only close channels on the RF MUX
page
179. For each path, the module stores the cycle count
page
179.