AC Voltage, AC Current (for all ranges):
AC Filter
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
Frequency, Period:
AC Filter
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
Digital Input, Totalize:
Channel Delay
0 seconds
Front Panel Operation: Channel (Configure) > CHANNEL DELAY > AUTO
Once you have added the specified channel to the scan list, the channel
delay choice will be visible in the menu.
Interface Operation: The following command enables an automatic channel
delay on the specified channels.
Selecting a specific channel delay using the
command (see
channel delay.
34980A User's Guide
Channel Delay
7.0 seconds
1.0 second
120 ms
Channel Delay
600 ms
300 ms
100 ms
ROUT:CHAN:DELAY:AUTO ON,(@1003,1013)
"Channel Delay"
ROUTe:CHANnel:DELay
on page 53) disables the automatic