Chapter 2. Verification Tests; Introduction; Test Conditions And Procedures; Performance Test Record - Agilent Technologies 75000 C Series Service Manual

32-channel, 5-amp, form c switch
Hide thumbs Also See for 75000 C Series:
Table of Contents

Advertisement

Introduction

Test Conditions
and Procedures
Performance Test
Record
Verification Test
Examples
Chapter 2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The three levels of test procedures described in this chapter are used to
verify that the Agilent E1463A 32-Channel Form C Switch Module:
is fully functional (Functional Verification)
meets selected testable specifications (Operation Verification)
meets all testable specifications (Performance Verification)
See Table 1-1 for test equipment requirements. You should complete the
Performance Verification tests at least once a year. For heavy use or severe
operating environments, perform the tests more often.
The verification tests assume that the person performing the tests
understands how to operate the mainframe, the switch module, and specified
test equipment. The test procedures do not specify equipment settings for
test equipment, except in general terms. It is assumed that a qualified,
service-trained technician will select and connect the cables, adapters, and
probes required for the test.
It is assumed that the temperature is no greater than 25
humidity is no greater than 40%.
The results of each Performance Verification test may be recorded in the
Performance Test Record (Table 2-1).
Each verification test procedure includes an example program that performs
the test. All example programs assume the following:
Controller is an HP 9000 Series 200/300 computer
Programming language is BASIC
Switch address is 70915
Switch card number is 1
Chapter 2
Verification Tests
o
Verification Tests 19
C and the relative

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1463a

Table of Contents