Display HOLD
To prevent electric shock, when Display HOLD is
activated, the display will not change when a
different current is applied.
Pressing activates Display Hold mode. The display shows
and the Meter freezes the display. To exit and return to
normal operation, press a second time.
Auto Off
The Meter automatically turns off after 15 minutes of inactivity. To
disable the Auto Off feature, hold down while turning on the
Meter. If the Meter has automatically shut down, restart it by
pushing ("WAKE UP").
Taking Measurements
The Clamp Meter is not for use on non-insulated
conductors.
Measurements can be taken with the clamp in the docked
position, or remotely via the 1 m cable. For accurate
measurements:
Always zero the Meter prior to taking measurements.
•
Zero the Meter as close to the measurement source as
•
possible.
Make sure the clamp is free of contamination.
•
To reduce magnetic influences, zero the Meter in the
same position or jaw direction that is used for the
measurements.