DOS & DON'TS
¬ Hold syringe vertically when filling disks.
¬ Do not fill disk while in the meter. Fill disk on clean, dry surface.
¬ The disk should not contain any large air bubbles. Air bubbles will
result in erroneous results.
¬ Only the Universal Disk Cover (Code 1719) can be used with the
WaterLink
Spin Touch
®
¬ Empty syringe of previous sample before filling with next
sample.
¬ Remove filled disk from meter after testing.
¬ Store disks at 70°-80°F/21°-27°C.
¬ Keep the chamber clean and dry. Gently swab LED and
photodiode lenses located around the hub with a cotton swab
dampened with streak-free window cleaner. Do not use alcohol.
It will leave a thin film over the lenses when dry.
FILLING: Multiple-Use Disks
When the syringe is placed in the water
sample, and the plunger is pulled all the
way up, the syringe will hold more than
enough water sample to adequately fill
the disk. Hold the syringe vertically and
insert the tip into the fill hole in the disk.
Press the plunger slowly and smoothly to
fill the disk.
2. Do not overfill the disk. If the disk is
overfilled, sample water will flow out
of the vent hole in the center of the
disk. The disk is not leaking. Dry the
disk and run the test.
4. Do not introduce air bubbles into
the disk. The reagent chambers
will not fill entirely and results will
be inaccurate. As soon as a bubble
starts to form, pull back on the
plunger to draw the bubble out of the
disk. Begin the filling process again.
6. Disks should be filled and used within 10 minutes. They cannot be filled
ahead of time.
Compatible with iPhone 6 Plus, iPhone 6, iPhone 5s,
iPhone 5c, iPhone 5, iPad Air 2, iPad Air, iPad mini 3,
iPad mini 2, iPad mini, and iPad (4th generation).
Compatible with Android devices.
.
®
Over Flow
Chamber
Blank
Chamber
Overfilled
(Dry disk before placing in meter)
For filling and troubleshooting tips go to:
See the WaterLink
410-778-3100 | 800-344-3100 | www.lamotte.com
LaMotte, WaterLink, Spin Touch are ® registered trademarks of LaMotte Company | © 2021 LaMotte Company, All Rights Reserved
CONNECTING VIA BLUETOOTH
The WaterLink
Bluetooth enabled device, such as a phone or tablet. The Spin
Touch is also compatible with a LaMotte Bluetooth Printer [Code
5-0066]. Other Bluetooth printers are not supported.
Transferring results over Bluetooth to a mobile device requires
that a mobile app is installed on the device and an active
account for the associated software program. For example, with
an active WaterLink
Solutions
transferred from the meter into the application. Mobile apps for
LaMotte software products are available from iTunes (for iOS
devices) and Google Play (for Android devices).
For detailed information read the Spin Touch manual. Go to
http://softwarecenter.lamotte.com/
LaMotte software products.
1. Fill any unused section of the disk
using slow, even pressure. Add
the sample water until the Blank
Chamber is full and the Over Flow
Chamber is partially filled.
3. Do not under fill the disk. If the disk
is under filled, the blank chamber will
not be full. The reagent chambers
will not fill entirely and results will be
inaccurate.
5. Wet disks should be dried thoroughly
with a lint free wipe. Handle the disk
by the edges. Between tests store
the disk on a flat surface. Do not
agitate a partially filled disk to avoid
leaking and contamination. Partially
used disks should be discarded at
the end of the day.
Air Bubble
www.waterlinkspintouch.com
Spin Touch
manual for additional information.
®
®
softwaresupport@lamotte.com
Spin Touch
is capable of connecting to a
®
®
Solutions
account and the WaterLink
®
mobile app on a phone or tablet, results can be
®
to learn more about
Over Flow
Chamber
Blank
Chamber
Correct Fill
Blank
Chamber
Underfilled
Disk US Patent No. 8,734,734
FCl US Patent No. 8,987,000
FCl EU Patent No. EP2784503 A1
TCl US Patent No. 8,993,337
3580-QG | 02.24.21