BEFORE FIRST USE
¬ Fully charge battery using
the Spin Touch USB cable and
adapter plug in AC outlet.
¬ Ensure your Spin Touch has the
latest fi rmware:
1. Download and install
®
the WaterLink
application for Windows
at
http://softwarecenter.
lamotte.com/.
2. Plug meter into the computer
with the provided USB cable
and launch WaterLink
Connect application from the
Start Menu.
3. Any fi rmware updates will
automatically be applied to
the meter.
FILLING: Single-Use Disks
1. Fill the disk using slow,
even pressure. The sample
water will fi ll the spaces
between the baffl es in a
counterclockwise order.
Each space will fi ll from the
bottom to the top. Sample
water should be added
until the sample water in
the fourth chamber fi lls
to the top of the chamber
slightly past the embossed
fi ll line. It is OK to fi ll
slightly past the fi ll line.
3. Do not under fi ll the disk. If
the disk is under fi lled, the
reagent chambers will not
fi ll entirely and results will
be inaccurate.
5. Wet disks should be dried
thoroughly with a lint free
wipe. The disk should be
handled by the edges.
TESTING
1. Press
2. Tap
3. Tap
confi rm.
4. Remove disk from packaging.
5. Use syringe (1189) to fi ll disk with pool or spa water sample.
Connect
6. Insert disk. Cover disk with Universal Disk Cover. Close lid.
®
7. Tap
8. Results will be displayed.
9. Select an option.
- Tap
- Tap highlighted
®
- Tap the hightlighted
- Tap highlighted
10. Press and hold
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the
most accurate result, wait until water sample is below 90 °F (32 °C) to test.
Correct Fill
(Fill line)
Under
Filled
For fi lling and troubleshooting tips go to:
QUICK
START GUIDE
and hold until the meter turns on.
. Select a sanitizer type. Tap
. Select a disk series (found on disk packaging). Tap
to start test. Tap
to return to the Test Screen.
to save test results to test log if Auto Save is
not enabled.
to send the results to the enabled
Mobile Bluetooth Printer.
to send results to Bluetooth enabled device.
for 2 seconds to turn meter off .
2. Do not overfi ll the disk.
If the disk is overfi lled,
sample water will fl ow
out of the overfl ow hole
in the center of the disk.
The disk is not leaking.
Dry the disk and run the
test.
4. Do not introduce air
bubbles into the disk.
The reagent chambers
will not fi ll entirely
and results will be
inaccurate. As soon as
a bubble starts to form,
pull back on the plunger
to draw the bubble out
of the disk. Begin the
fi lling process again.
6. Disks should be fi lled and used within 10 minutes.
They cannot be fi lled ahead of time.
www.lamotte.com/spin
to confi rm.
to cancel the test.
Over
Filled
(Dry disk before placing in meter)
Air
Bubble
to
Need help?
Do you have a question about the WaterLink Spin Touch and is the answer not in the manual?