Safety notes
2.4 Installation in hazardous locations
• Sensor insulation thickness is max. 100mm (only insulated sensors).
• IEC/EN 61241-14 and 61241-17 are considered for installation in areas with combustible
dust.
16
DANGER
Risk of explosion
Do NOT open the device while energized; otherwise there is a risk of explosion.
WARNING
Explosive gas or dust
The terminal box shall not be opened when an explosive gas or dust atmosphere may be
present.
WARNING
Potential equalization
In operation, the output is earthed through the conductive medium being measured and
therefore potential equalization is necessary throughout the hazardous area.
The apparatus housing shall be connected to the potential equalizing conductor in the
hazardous area.
WARNING
External connections to Ex e terminals
The external connections to the Ex 'e' Terminals of the Remote version shall comply with the
following:
• The wire conductors shall have a cross-sectional area between 0.5 mm
• No more than one single or multiple strand wire conductor shall normally be connected
to each of the terminals. If multiple conductors are required, these shall be joined in a
suitable manner, for example two conductors into a single insulated crimped boot lace
ferrule.
• The insulation on the wire conductors shall extend to within 1 mm of the metal of the
terminal throat.
• The terminal screws shall be tightened down with a torque between 0.5 Nm and 0.7 Nm.
• The terminals shall never be exposed to temperatures outside of the range -50°C to +
130°C; in addition, they shall only be installed and wired with cable in an ambient
temperature of -10 to +80°C. Furthermore, in the event of there being a process
temperatures of + 180°C in conjunction with an upper ambient temperature of +50°C
the terminal strips should not be installed or wired with cable.
SITRANS FM MAG 3100
Operating Instructions, 12/2021, A5E03005599-AE
2
and 2.5 mm
2
.