I.i. Laser Light Localizer; Dicom Bridge; Codonics Printer - Siemens SIREMOBIL Iso-C Maintenance Instruction

Table of Contents

Advertisement

30
3.1.1

I.I. laser light localizer

SIM
Mechanical system
Inspect the I.I. laser light localizer for mechanical damage.
Mount the I.I. laser light localizer on the I.I. and ensure proper locking and seating.
When doing this, pay special attention to the tension band and its closure.
Check the I.I. ring for damage.
SIM
Function
Perform maintenance on the I.I. laser light localizer according to Installation and Set-
ting Instructions RXR2-130.033.03...
3.1.2
Laser light localizer
SIM
Mechanical system
Inspect the I.I. cap of the laser light localizer for mechanical damage.
Install the I.I. cap of the laser light localizer on the I.I. and ensure proper locking and
seating.
Check the I.I. ring for damage.
SIM
Function
Perform the maintenance steps specified according to the adjustment instructions
SPR2-230.032.01..
3.1.3

DICOM bridge

SIM
Mounting
Check the mounting of the DICOM bridge in the monitor trolley.
3.1.4

Codonics printer

SIM
Mounting
Inspect the screws used to fasten guide rail to the mounting bracket of the Framegrab-
ber.
Ensure that the printer is properly fastened to the guide rail. Make sure there are no
missing attachment screws.
Ensure that the Framegrabber is properly fastened to the mounting bracket. Make sure
there are no missing attachment screws.
SIM
Guide rails
Inspect the guide rails for proper function and smooth operation.
The guide rails must not be damaged.
The guide rails must lock properly into place when fully pushed into position.
If necessary, use a cleaning brush to remove dirt from the rails.
SIREMOBIL Iso-C
SPR2-230.101.02.06.02
07.05
Safety Inspection
Page 30 of 46
CS PS 24
Siemens AG
Medical Solutions