Safety Inspection
3-
3
Safety Inspection
3.1

Mechanical safety

SIM
Cover panels
Remove the covers from the SIREMOBIL stand and monitor trolley.
Check cover panels for damage.
SIM
Cable deflectors
Inspect the cable deflectors on the stand and on the monitor trolley.
SIM
Foot brake
Test the function of the foot brakes on the SIREMOBIL stand and monitor trolley on a
level floor.
SIM
Brakes
Use the spring balance to test whether the orbital, angulation, swivel and horizontal
movement brakes apply the prescribed force. (Refer to service instructions SIREMO-
BIL ISO-C)
SIM
C-arm
Check the C-arm for damage to the carbon fiber structures (visual inspection). If there is
damage, proceed as described in the warning in chapter 1.
Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
Orbital movement: Once the orbital brake is released, the C-arm can be moved only in a
limited manner. Check whether the orbital movement is smooth.
SIM
Wheels
Move the SIREMOBIL in a straight line on a level floor.
Evaluate the straight and quiet movement of the SIREMOBIL.
Replace any defective wheels or castors.
Check the steering function.
SIM
Lifting column
Switch the system on
Electrically raise and lower the lifting column,
- while listening for noises and checking for play in the bearings.
- The lifting column movement must switch off automatically when the end positions
are reached.
Switch off the system.
Lubricate the lifting column.
- Lubricate the bearing surfaces of the lifting column guide and the spindle.
SIM
Emergency Stop switch
Press the Emergency Stop switch. The lifting column may not be raised or lowered.
Press the Emergency Stop switch again. The lifting column can be moved again.