SIREMOBIL Iso-C
Installation- and Setting Instructions
Area dose product measurement device
Valid for
SIREMOBIL Iso-C
and
SIREMOBIL Compact L from Serialnumber >3000
Print No.: SPR2-230.033.04.03.02
Replaces: SPR2-230.033.04.02.02
SP
© Siemens AG 1998
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 06.05

Advertisement

loading

Summary of Contents for Siemens SIREMOBIL Iso-C SP

  • Page 1 SIREMOBIL Iso-C Installation- and Setting Instructions Area dose product measurement device Valid for © Siemens AG 1998 SIREMOBIL Iso-C The reproduction, transmission or use of this document or its contents is not permitted without express written SIREMOBIL Compact L from Serialnumber >3000 authority.
  • Page 2 Center cross e.g. 96 60 051 RE999 Silicon sealing compound 20 49 716 Tools required Standard service kit 1 set of Allen keys SIREMOBIL Iso-C SPR2-230.033.04 Page 2 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 3: Safety Information

    TD00-000.860.01... must be com- plied with. CAUTION Electrical voltage! Non-compliance can lead to material damage. When working on the system, ESD regulations must be observed. Siemens AG SPR2-230.033.04 Page 3 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 03 06.05...
  • Page 4 Perform the particular work steps with special care and attention to detail. If needed, wear working gloves. SIREMOBIL Iso-C SPR2-230.033.04 Page 4 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 5 Avoid all contact with blood or other bodily fluids! Strictly observe the safety information in ARTD-002.731.37... regarding prevention of infectious diseases during customer ser- vice calls. Siemens AG SPR2-230.033.04 Page 5 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 03 06.05...
  • Page 6 - Release radiation for as short a time as possible. Checks in which radiation must be released are identified by the radiation warning symbol. SIREMOBIL Iso-C SPR2-230.033.04 Page 6 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 7 (e.g. monitor cable between basic unit and monitor trolley) must be moved section by section to detect broken conductors. The protective conductor resistance may not exceed 0.2 Ohms. Siemens AG SPR2-230.033.04 Page 7 of 16 SIREMOBIL Iso-C Medical Engineering Rev.
  • Page 8 DIN VDE 0751-1:2001-10, Fig. C2. 1 = System 2 = Application part (not present) 3 = Measuring arrangement (integrated in the measuring instrument) SIREMOBIL Iso-C SPR2-230.033.04 Page 8 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 9: Installation

    Check the laser targeting device according to the adjustment instructions and, if necessary, adjust it. • Refit the SIREPHOS cover hood and seal the joints with silicone sealing compound. Siemens AG SPR2-230.033.04 Page 9 of 16 SIREMOBIL Iso-C Medical Engineering Rev.
  • Page 10 Select the sub-menu "Save" and save the parameters on disk. • Exit the Memoskop service program. • Turn the SIREMOBIL off and the on again to reset the serial interface. SIREMOBIL Iso-C SPR2-230.033.04 Page 10 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 11 • Click on "Save to File". You must exit the service program with "Logoff" and "Quit" to NOTE store the modified parameters to disk as a new file. Siemens AG SPR2-230.033.04 Page 11 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 03 06.05...
  • Page 12 The potentiometer (fig.2) of the area dose product measurement device can be used to adjust the test value to the programmed value (Qt*Kg). SIREMOBIL Iso-C SPR2-230.033.04 Page 12 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 13 After the adjustment, select "Stop Diamentor Test" on the service PC and click on "Execute". • The test is ended. • Check the accuracy of the area dose product measurement device. Refer to the next page. Siemens AG SPR2-230.033.04 Page 13 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 03 06.05...
  • Page 14 Checking the accuracy of the area dose product measurement device Prerequisites The calibration of the area dose product measurement device has already been per- formed. SIREMOBIL Iso-C SPR2-230.033.04 Page 14 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...
  • Page 15 Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ] / (measured area dose product FDPg) For deviations > 0.3 , the area dose product measurement device must be replaced. Siemens AG SPR2-230.033.04 Page 15 of 16...
  • Page 16: Concluding Steps

    Required measuring equipment and auxillary devices revised. Check of the programmed values revised. Check of the accuracy of the area dose product measuring instrument revised. SIREMOBIL Iso-C SPR2-230.033.04 Page 16 of 16 Siemens AG Rev. 03 06.05 CS PS 24 Medical Engineering...