Series 5000TC Toroidal (electrodeless) Conductivity Sensors
3.4

Insertion/Removal Sensors Mounting Types

Location Requirements
Mount the sensor vertically with electrodes pointing downward. If the sensor must be installed on an angle,
it should be at least 15° above horizontal. Other mounting angles may cause erratic measurement readings.
Mounting
When following the instructions in this section, refer to Figure 3-5 or Figure 3-6. The instructions refer to
specific component parts that are identified by letter designations in the respective figures.
ATTENTION
Use Teflon tape on all sensor and mounting hardware threads to avoid leaks. (If using pipe sealant with
Teflon, Locktite No. 59321 or equivalent, it may not provide an adequate seal at higher solution
temperatures.)
12
Series 5000TC Toroidal (electrodeless) Conductivity Sensors - Operation and Maintenance

Figure 3-4 Sanitary (CIP) Flange Mounting Details

4/00