Table 4-2 Termination Procedure For Ce-Only Applications; Figure 4-2 Ce Termination Details (For Ce-Only Applications) - Honeywell 5000TC Series Operation And Maintenance Manual

Toroidal (electrodeless) conductivity sensors
Table of Contents

Advertisement

CE Termination Procedure (For CE-only Applications)
CAUTION
When the measuring instrument has the European community "CE" label, this CE termination procedure
must be used. If there is no "CE" label on the instrument, use the standard termination procedure.
Step
1
Carefully strip back the outer cable jacket and outer foil wrap to an appropriate length. This exposes
the signal shield wire, cable shield wire, and cellophane-wrapped wire bundle.
2
Fold back the cable shield wire (see Figure 4-2). Using shrink wrap or tape, equally cover the outer
cable jacket, and the exposed cable shield wire/cellophane-wrapped wire bundle.
3
Carefully clip back the cellophane wrap, inner foil wrap, and unused black wire up to the end of the
shrink wrap or tape applied in step 2. This exposes the signal shield wire, and the five colored wires
(red, green, yellow, white, and blue).
4
Insulate the signal shield wire with shrink wrap or tape.
5
Strip the insulation back from each of the colored wires to a distance of 1/8 inch (3 mm). Then tin each
stripped wire.
6
Using an ohmmeter or test light, verify that the signal shield wire is not shorted to the cable shield wire. If
okay, shorten the cable shield wire to an appropriate length. If not, clip cable to get a new unfinished end
and start over at step 1.
7
Using shrink wrap or tape, equally overlap the existing shrink wrap or tape applied in step 2, and the five
colored wires/signal shield wire. This secures and completes the terminated assembly
4/00
Series 5000TC Toroidal (electrodeless) Conductivity Sensors - Operation and Maintenance

Table 4-2 Termination Procedure for CE-only Applications

867G@ÃTCD@G9ÃXDS@
76S@ÃXDS@
TDBI6GÃTCD@G9ÃXDS@
PVU@SÃ867G@
E68F@U

Figure 4-2 CE Termination Details (for CE-only Applications)

Sensor/Interconnect Cable Termination
Action
TCSDIFÃXS6QQ@9ÃPSÃU6Q@9
TDBI6GÃTCD@G9ÃXDS@
DII@SÃAPDGÃXS6Q
8@GGPQC6I@ÃXS6Q
PVU@SÃAPDGÃXS6Q
6QQSPQSD6U@ÃG@IBUC
APSÃ8PII@8UDPI
TCSDIFÃXS6QÃPSÃU6Q@
'ÃDI8C
"À€