MAN0364.P65 Issue 03, Mar 01
00705M5003
TP3
2.5 FITTING SENSORS
2. INSTALLATION
Connection Terminals
Offset
Circuit Board Assembly Showing
Adjustment Controls
1.
Before installing sensors, isolate the power supply
by disconnecting or removing the associated control
module from the installation.
2.
All electrochemical cells contain acid or a caustic
solution. Avoid its contact with skin, eyes and clothing.
In case of contact flush with water for at least fifteen
minutes.
The sintered disc on the sensor assembly must be kept
free from contaminants, for example oil and dirt.
Span
TP2
Zero
Figure 7 Connections
WARNINGS
CAUTION
14
705 Sensor CO
RFI Shield Plate