Removing The; Removing The X-Ray Source Assembly - Siemens MULTIMOBIL 2.5 Service Manual

Hide thumbs Also See for MULTIMOBIL 2.5:
Table of Contents

Advertisement

8.10.2 Removing the Front cover.

8.10.3 Removing the X-ray source assembly.

Turn off the Multimobil 2.5,disconnect the mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to transport position. Lock the movement of the
Multimobil 2.5 by applying the foot brake.
Siemens Ltd. Med India
Turn off the Multimobil 2.5,disconnect the mains chord
and wind the cord around the cable holder.
Lower the X-ray source assembly to transport position.
Lock the Multimobil 2.5 by applying the brake.
Arrange some kind of soft bedding to place the front
cover on .
Unscrew the four ornamental screws with 2.5mm allen
key.
Grasp the cassette box and slowly pull it outside.
Position the front cover on a soft bedding.
Version 6.0