Control Solution Testing - Contour Next GEN 9658 User Manual

Table of Contents

Advertisement

CAUTION: Do not use expired materials.
Using expired material can cause inaccurate
results. Always check the expiration dates on
your test materials.
Level 1 and Level 2 control solutions are available and sold separately if
not included in the meter kit. You should test your C
meter with control solution only when the temperature is 59°F–95°F.
Always use C
n
control solutions. Other brands could yield
ontour
ext
incorrect results. If you need help locating C
solutions, contact Customer Service. See the back cover for contact
information.

Control Solution Testing

1. Remove a C
n
test strip from the bottle.
ontour
ext
2. Tightly close the bottle lid immediately after you remove the test strip.
3. Hold the test strip with the grey square end facing up.
26
n
GEN
ontour
ext
n
control
ontour
ext
4. Insert the grey square end of the
test strip into the test strip port
until the meter beeps.
The meter turns on, displaying a test strip with a
10 45
4 12 22
flashing blood drop.
CAUTION: Do not use control solution that is more than 6 months
past the date you first opened the bottle.
NOTE: If this is the first time you are opening the control solution,
write the date on the bottle.
5. Shake the control solution bottle well, about
15 times before every use.
6. Remove the bottle cap and use a tissue to wipe away any solution
around the bottle tip before dispensing a drop.
7. Squeeze a small drop of solution onto a clean, nonabsorbent surface.
CAUTION: Do not apply control solution to your fingertip or to the
test strip directly from the bottle.
C
n
GEN blood glucose monitoring system
ontour
®
ext
TESTING
27