Cleaning And Sterilization; Manual Wash - Integra MAYFIELD Swivel Adaptor Instruction Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
WARNING:
The unit should be checked to ensure that
it has been properly locked. In the locked position,
pressure applied in either direction to the headrest
WILL NOT result in movement of the Swivel
Adaptor.
Figure 4

Cleaning and Sterilization

After each use, thoroughly clean and wipe the unit
with an antiseptic solution.
DO NOT STEAM STERILIZE !
Plastic components may be damaged by
heat.

Manual Wash

CAUTIONS
• Alkaline and highly acidic detergents and
solutions cause damage to the devices.
• Channels and crevices found on this device
require particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can damage
the surface of the equipment. Avoid using
hard water. Instead use purified water unless
otherwise specified.
Limitations on reprocessing
• Repeated processing has minimal effects on these
devices. Product life is normally determined by
wear and damage due to use.
• It is important to have Integra NeuroSpecialists
perform routine inspections (twice yearly is
recommended). See contact information below.
INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for
safe containment and transport to the
decontamination environment.
• It is recommended that devices are cleaned
immediately after use.
Preparation for Cleaning
• Disassembly is required.
Disassembling
Equipment
Completely remove the
swivel adaptor from
any components. The
swivel adaptor itself
cannot be disassembled
any further.
Cleaning – Manual Equipment: Water, Neutral pH
Detergent, Soft Bristle Brush, Towels
Method
1. Prepare neutral pH enzymatic detergent solution
(e.g. Endozime® AW Triple Plus with APA (Ruhof),
1:128 ratio) according to detergent manufacturer's
instructions using lukewarm tap water.
2. Prepare equipment for soaking by disassembling
removable parts and loosening connections.
3. Rinse equipment in warm water before placing
into bath.
4. Completely soak equipment in water/detergent
solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush.
NOTE – If possible, use a disposable brush.
6. Rinse in warm purified water until all visible
substances and residual detergent are removed.
NOTE – Make sure to give special attention to
hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels
and use medically compressed air if needed, to
dry channels, crevices and lumens.
8. Inspect the equipment to make sure there is no
visible organic debris or residue from cleaning
agent.
Repeat process if any soil is detected.
7
EN – ENGLISH
Inspecting for
Cleanliness
Inspect the teeth of
the starbursts, the area
around the locking
screw, and the base of
the swivel adaptor for
organic debris.