Automatic oil acidity tester

Advertisement

Quick Links

HZCS-3
Automatic Oil Acidity Tester
Huazheng Electric Manufacturing (Baoding) Co., Ltd

Advertisement

loading
Need help?

Need help?

Do you have a question about the HZCS-3 and is the answer not in the manual?

Questions and answers

Summary of Contents for Huazheng HZCS-3

  • Page 1 HZCS-3 Automatic Oil Acidity Tester Huazheng Electric Manufacturing (Baoding) Co., Ltd...
  • Page 2: Table Of Contents

    Contents I. Overview and Technical Index....................2 II. Structure and Installation....................3 III. Preparation Work........................3 IV. Instrument Operation......................5 V. Maintenance........................21 VI. Operation Notice......................22 VII.Packing List........................23...
  • Page 3 Dear customers: Thank you for choosing HZCS-3 transformer oil acid tester. We hope that this instrument will make your work more relaxed and happy, So that you can get clear and accurate results in the experimental analysis work. Please read this User Manual before using the tester, and follow the instructions to operate and maintain it for prolonging its service life.
  • Page 4: Overview And Technical Index

    I. Overview and Technical Index Overview HZ CS-3 transformer oil acid tester is a fully aut omat ic measuring inst rum ent which adopt s the ex tr action method to accur ately det ect the insulating oil ac id value. This inst rum ent can impr ove the work ing eff iciency and test ac curacy, and at the same time ensur e the safet y of the body by reducing the oper at or's contact wit h the samples and reagents .
  • Page 5: Structure And Installation

    II. Structure and Installation Host structure Back Cover Front Cover Installation After unpacking, put the instrument in a stable laboratory bench. Then check whether the parts are damaged or not, and whether the random accessories are complete. Connect the power supply. Note:The instrument must be adequately grounded.
  • Page 6 alcohol and 500ml petroleum ether into the A container; ② Take 1 bottle of extract medicine block (see above) into reaction cup B, add 10ml isopropanol and 10ml petroleum ether, stir the extract medicine block, isopropanol and petroleum ether in the reaction cup, stir and dissolve into the pre In the prepared A container;...
  • Page 7: Instrument Operation

    (7)Anhydrous ethanol:Inject Anhydrous Ethanol in the wash bottle provided with the instrument. Pipeline connection Open the instrument back cover,put in the three plastic bottles of anhydrous ethanol, washing gas cylinders, neutralizing liquid (ethanol-KOH) from left to right. Connect another port of neutralizing liquid (ethanol-KOH) to the interface affixed with a neutral label with a hose;...
  • Page 8 If there is no sample cup on the sample tray, the instrument enters the following interface: Open the front cover of the instrument and place it in the sample cup (placed in the second cup position). After the sample cup is detected, the instrument starts to extract the extract.
  • Page 9 When there is 20 drops, the instrument enters the main interface: Click the time shown on the upper right corner, the instrument enters the following interface and time can reset.
  • Page 10 Input the date and time in turn and click the “OK” button,time can reset successfully. Note: Stirring should be placed in the cup for calibration and sample testing. Instrument Calibration If the instrument is used in the first time or the neutralization is being changed , this step is necessary.
  • Page 11 According the prompt, inject 20ml BTB aqueous solution accurately in a sample cup (see details in the chapter 3 BTB aqueous solution) and put the cup in the sample tray. Click the “Start” button and a blank titration is beginning. The instrument enters the following interface: When the blank titration is finished, if a re-calibration is in progress, the instrument enters the following interface:...
  • Page 12 If verification is in progress, the instrument enters the following interface: Input the volume of the standard acid solution; inject corresponding 0.15mol/L standard acid solution with micro injector (see details in the chapter 3 “Standard acid solution”) in the cup. Click the “Start” button, the neutralization is in progress, the instrument enters the following interface:...
  • Page 13 When the neutralization is finished, if a re-calibration is in progress, the instruments update the calibration value and enter the main calibration interface. If verification is in progress the instrument enters the following interface: Click the “Print” button to print the result of verification or go back the main calibration interface.
  • Page 14 If there is no enough extract, the instrument warns like the following interface: Add enough extract and press “OK”, after the self-inspected the instrument enters the following interface:...
  • Page 15 It should be noted that the extract is subjected to blank titration after the addition of the extract. The specific steps are to put a clean sample cup into the sample tray. Click the “Blank” button to start blank titration of the extract. The instrument enters the following interface: Looking for the sample cup, after this progress the interface will prompt to change the cup, the interface is as following:...
  • Page 16 Replace the sample cup with a clean sample cup. Click OK to start extracting 20ml of extract. Then blank titration of the extract is in progress, the interface is as following:...
  • Page 17 The number of blank titration drops is automatically saved after the titration is completed. The extract should be blank calibrated in the following cases: replace the extract; Replace the neutralizing solution. Then the instrument enters the following interface: Input the oil density, click the “Turn” button and put cups in turn, click the “Start” button and the instrument enters the following interface:...
  • Page 18 The instrument starts looking for the first cup, when the first cup is founded and the cup is founded, if there is no enough extract the instrument enters the following interface: After add enough extract, press the “OK”button to continue testing. The instrument enters the following interface:...
  • Page 19 When this step finished the extraction is back-flowing, the instrument enters the following interface: When this step finished the neutralization starts, the instrument enters the following interface:...
  • Page 20 When this step finished the result is shown and the instrument is looking for the next cup, the instrument enters the following interface: If there is no sample cup, the instrument is looking for the next cup and shows as the following interface:...
  • Page 21 When this test is finished, the instrument enters the “History data query” interface (see details in the part below) and the operator shall take out all cups and clean up it before the next test or closing the instrument. History data query In the main interface click the “History Data Query”...
  • Page 22 the following interface. If there is no history data, the instrument enters the following interface: Instrument cleaning Put in a sample cup and press the “Instrument Cleaning” button in the main interface, the instrument starts cleaning up and enters the following interface:...
  • Page 23: Maintenance

    Turn off the power of the instrument. Take out the cup and clean up it. Open the back cover of the instrument then loosen the hose arm of the upper part of the peristaltic pump, then close the back cover. Maintenance Replace the reagent Replace a bottle of anhydrous ethanol when it run out ;...
  • Page 24: Operation Notice

    BUTTON Open the button shown on the picture above, and put in a new printer paper and close the cover of printer. If there is no word, check whether the printing paper is reversed。 Shut down the power and press the “LF” button continuously, at the same time turn on the power, after 5s release the “LF”...
  • Page 25: Vii.packing List

    VII.Packing List Item Main engine Neutralization liquid medicine Extracted liquid medicine Gas washing bottle medicine Indicator Standard acid Microsyringe Sample cup Stirrer Glass bottle Glass bottle cap Tubing for Peristaltic Pumps Print paper Power line...