Summary of Contents for Huazheng HZJY-10K-Z

  • Page 1 HZJY-10K-Z Insulation Resistance Tester Huazheng Electric Manufacturing (Baoding) Co., Ltd...
  • Page 2 (0312) 6775656 to tell you to ser ve you at all times-B aoding Huazheng Elect ric Manuf act uring Co., Lt d. , our company will def initely make you satisf ied !
  • Page 3: Table Of Contents

    Contents I.Safe use items ..........................2 II.Feature ............................3 III.Specification Parameter ......................4 IV.Introduction of instrument .......................5 V.Instrument using ........................8 VI. Appendix ..........................9...
  • Page 4: I.safe Use Items

    I.Safe use items Design and manufacture of the instrument and testing achieves IEC61010 safety standards (electronic measuring product safety requirements), this manual including the safety of the instrument use and ensure the safety of equipment state, users must follow the warnings and safety regulations Please read the following instructions before use.
  • Page 5: Ii.feature

    ● Make sure all wires and instrument testing port connection is firm. Note (for improper operation may cause personal injury or equipment damage) Squared before measurement, confirm the voltage selection in the appropriate value. If long time not to use, the battery should be fully charge, and recharge in 3 months. Squared do not in high temperature, damp, places and condensation may be placed under direct sunlight for a long time.
  • Page 6: Iii.specification Parameter

    ● digital filtering function. When the displayed value deviates due to external influence, the filtering function is used to reduce the influence. ● perfect protection function and prompt function for fuse burnout. ● data saving function and data export. III.Specification Parameter ●...
  • Page 7: Iv.introduction Of Instrument

    temperature: 45 – 75% RH) range 500V 1000V 2500V 5000V 10000V Other voltages accuracy Not specified <100K <100K <100K <100K <100K Range: 1G/ V, 100G at ±5%rdg±3dgt 100K -100G 100K-200G 100K-500G 100K-1T 100K-2T 100V. When the voltage is less than 200V, the ±20%rdg 100G-1T 200G-2T...
  • Page 8 key number Description Press the switch on button startup, shutdown condition 2、3、4 functional key The TEST key, press and hold 1 s TEST, then stop (enter the Settings menu is invalid) ● Test setup interface Test voltage Test duration Battery setting level 10:00...
  • Page 9 the insulation resistance value under 5 voltages. Key name Description Set the current value in case of short circuit, increasing by Set the current value in case of short circuit, decreasing by Set the filter: 0:without filtering; (no interference case recommended) 1:hardware low-pass filter;...
  • Page 10: V.instrument Using

    HELP :Record the serial number of product, production date, version information. ● Insulation test interface Actual test Insulation Measured battery voltage Simulated time level 10:00 5000V Leakage current Set HV value ∞ 100k 1M 10M 100M 1G 100G 1T 10T 1mA=1000u High voltage DD timing,Display...
  • Page 11: Vi. Appendix

    current value and the capacitance value of the tested object 1 minute after the test is completed. DD value = current value 1 minute after test (mA) / voltage value at test completion (V) / capacitance value (F) DD value 2.0 less 2.0 ~...