3. Get started with Emfit Epilepsy Alarm
NOTE:
Always test the device before using it and when the settings are adjusted.
3.1. Set up the DIP-switches before use
Inside, you will find eight DIP switches that are used to select the desired functions (see picture).
Default factory settings
Movement notification delay is 13 seconds.
The bed exit notification is disabled.
The SW1 switch acts as the On/Off switch, which means that
the device can be switched on or off by pressing the switch
for three (3) seconds.
The notification sound is set to Very loud.
NOTE:
If the wrong DIP switch settings are used, the unit will not work as expected.
•
Remove the batteries and disconnect the power supply before setting up the DIP
•
switches to implement the new settings.
To change the settings:
1. Disconnect the power supply (if used) and all cables.
2. Open the lid by lifting the side, see image.
3. Remove the batteries.
NOTE:
If the batteries and power supply are not disconnected, the
new settings will not be implemented.
4. Select the desired settings.
5. Insert the batteries.
6. Close the lid
7. Connect all cables and the power supply.
8
Switch
1 and 2 OFF
(down)
3 ON (up)
4 and 5 OFF
(down)
6 OFF (down)
7 and 8 OFF
(down)