4.2.3. Use SW1 as a bed exit alarm bypass switch
The bed exit alarm can be deactivated if the user wants to leave the bed without triggering an
alarm. This means that the user can leave the bed without any time limit. The time delay for the
bed exit alarm is not activated again until the user returns to the bed and then leaves it.
When the bed exit alarm is activated through DIP switch 4 and you want to deactivate the alarm,
the SW1 switch must be pressed before the time delay runs out. A beep sound is heard when the
alarm is deactivated. The SW1 switch can also be pressed before the user leaves the bed to
deactivate the alarm.
When the SW1 switch is pressed, the user has 20 seconds to leave the bed. If not, the alarm is
activated again.
4.2.4. Use SW1 as a bed exit alarm "presence/absence" sensitivity switch
The SW1 switch can be used to launch automatic calibration for presence and absence sensitivity.
See chapter 4.5. Calibrating the "presence / absence" sensitivity

4.3. Signal lights

The LED lights provide information about the status of the product.
4.3.1. Green – Presence
Light flashes every second time
compared to blue light.
Light flashes the same speed as blue
light.
Light flashes rapidly.
Light is off.
4.3.2. Blue – Device on/standby
Light is off.
Light flashes slowly.
Light flashes rapidly a few times.
4.3.3. Red – Malfunction
Light is flashing the same speed as
blue light
Light is flashing every fourth time
compared to blue light.
14
A person is on the bed or other movement is being
made.
A person has been on the bed for 60 seconds and the
device has become activated.
Fast-paced movement on the bed detected
No one is on the bed.
Device is switched off
Device is switched on.
Device triggers a notification.
Bed sensor is disconnected or defective. Signal alarm
sounds after 10 seconds and then every 45 seconds.
There is also an alarm sent through the X2 connector
after 30 seconds and then every 30 minutes.
Batteries are empty. Replace batteries.