10. EMC INFORMATION
Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the device should assure that is used in such an environment.
Emissions
Compliance
RF emissions CISPR 11
Group 1
RP emissions CISPR 11
Class B
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/ flicker
Complies
emissions IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equip-
ment and the IH25/1
The IH25/1 is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the IH25/1 can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the IH25/1 as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter m
Rated maximum
output power of
150 kHz to 80 MHz
transmitter W
d = 1.2√P
0.01
0.12
0.1
0.38
1
1.2
10
3,8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Notes:
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
Electromagnetic environment—guidance
The device uses RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and not likely to cause
any interference in nearby electronic
equipment.
The device is suitable for use in all estab-
lishments, including domestic establish-
ments and those directly connected to the
public low-voltage power supply network
that supplies buildings used for domestic
purposes.
80 to 800 MHz
800 MHz to 2,5 GHz
d = 1.2√P
d = 2.3√P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
USER'S MANUAL
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an environment.
Immunity test
EN 60601 test level
Electrostatic
±6 KV contact
discharge (ESD)
±8 KV air
IEC 61000-4-2
Electrical fast t
±2 KV for power supply
transient/burst
lines
IEC 61000-4-4
Surge IEC 61000-
±1 KV line(s) and neutral
4-5
Voltage dips,
5 % UT
short
(>95 % dip in UT)
interruptions
for 0.5 cycle
and voltage
40 % UT
variations on
(60 % dip in UT)
power supply
for 5 cycles
input lines
70 % UT
IEC 61000-4-11
(30 % dip in UT)
for 25 cycles
>5 % UT
(>95 % dip in UT)
for 5s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
29
Electromagnetic environ-
Compliance level
ment - guidance
±6 KV contact
Floors should be wood,
±8 KV air
concrete or tile. If floors are
covered with synthetic ma-
terial, the relative humidity
should be at least 30%.
±2 KV for power supply
Mains power quality should
lines
be that of a typical com-
mercial or hospital envi-
ronment.
±1 KV line(s) and neutral Mains power quality should
be that of a typical com-
mercial or hospital envi-
ronment.
<5 % UT
Mains power quality should
(>95 % dip in UT)
be that of a typical com-
for 0.5 cycle
mercial or hospital envi-
40 % UT
ronment. If the user of the
(60 % dip in UT)
device requires continued
for 5 cycles
operation
during
mains interruptions, it is
70 % UT
(30 % dip in UT)
recommended that the de-
for 25 cycles
vice be powered from an
>5 % UT
uninterruptible power sup-
(>95 % dip in UT)
ply or a battery.
for 5s
3 A/m
Mains power quality should
be that of a typical commer-
cial or hospital
environment.