Initial Operation
Setting to Work
6.2.2 Cooling Circuit Connection and Filling
CAUTION
Risk of damage by using improper cooling hoses!
This may lead to damage to persons, damage to the unit or corrosion damage.
• When choosing cooling hoses pay attention to sufficient burst strength and compatibility with coolant!
• Only use cooling hoses without any signs of damage!
The cooling hoses with an internal diameter of 9 mm are connected to the unit by means of hose nipples.
Liquid outlet and liquid inlet are indicated with respective symbols.
Fig. 8:
Labeling of liquid inlet and liquid outlet
1)
Connect a suitable hose to the hose nipples for liquid inlet and liquid outlet and secure it with a clamp
respectively.
2)
Connect the hoses to the corresponding nipples of the device to be cooled.
PLEASE NOTE
When connecting the cooling hoses pay attention to flow direction. Follow the documentation released by
the manufacturer of the device to be cooled.
Transparent hoses stimulate algae growth that may increase the error-proneness of the components built
into the unit. Thus only use non-transparent hoses.
Fig. 9:
Cap of coolant container
1 Cap
Laird Thermal Systems s.r.o
20
Operation Manual, 387009851
Liquid outlet
Liquid inlet
Date: 08/09/2022
Version: 1.0
1
Liquid-Air Cooling Unit LA 5000
387009381