Laird LA 5000 Operation Manual

Liquid-air cooling unit
Table of Contents

Advertisement

Quick Links

Operation Manual
Liquid-Air Cooling Unit LA 5000
Version 1.0

Advertisement

Table of Contents
loading

Summary of Contents for Laird LA 5000

  • Page 1 Operation Manual Liquid-Air Cooling Unit LA 5000 Version 1.0...
  • Page 2 Laird Thermal Systems s.r.o. Prumyslová 497 462 11 Liberec Czech Republic Phone: +420 488 575 111 Protection Notice Dissemination as well as copying of this document, utilization or posting of its contents is not permitted unless formally granted. Violations entail claims for compensation. All copyrights, including rights due to patent grant or registration of a utility model or design, are reserved.
  • Page 3: Table Of Contents

    4.5.3 Infrastructure ..............................17 5 Transport ............................18 5.1 Safety Indications for Transportation and Setting-up ............18 5.2 Transportation of the Unit ......................18 Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 4 11.2 Drain of unit ............................. 33 11.3 Re-packaging of the unit ......................33 11.4 Storing the Unit ..........................33 11.5 Disposal ............................33 Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 5 Table of Contents 11.6 Disposal of Operating Materials....................33 11.7 Return of the unit to LAIRD THERMAL SYSTEMS ............. 34 12 Wear Parts and Spare Parts ....................35 12.1 Units wear and spare parts ......................35 Addendum ............................37 Flow scheme ............................. 37 Wiring diagram ............................
  • Page 6: About This Manual

    Any form of unintended use of the unit and any structural change caused by the user without prior authorization in written form by LAIRD THERMAL SYSTEMS will lead to the termination of warranty as well the termination of the declaration of conformation and will free LAIRD THERMAL SYSTEMS from product liability.
  • Page 7: Contact Information

    Post: Laird Thermal Systems s.r.o. Prumyslová 497 462 11 Liberec Czech Republic Tel: +420 488 575 111 Fax: +420 488 575 303 E-Mail: info-lcs@lairdthermal.com Homepage: www.lairdthermal.com Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 8: Product Identification

    1 Identification plate Fig. 2: Unit specific identification plate 1 Unit type 2 Article number 3 Serial number 4 Electrical specifications 5 Date of manufacture Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 9: Safety Regulations

    Warning with respect to electrical hazard Table 2: Warning signs Symbol Meaning This symbol indicates that disconnecting from mains is required. Table 3: Signs giving orders Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 10: Hints For Safe Operation

    All intended changes must be authorized by the manufacturer. • After the implementation of a change the operativeness of the safeguards must be verified. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 11: Environmental Issues

    • The minimum coolant level in the tank is secured by a level sensor. 1 Safety valve 2 Thermostat 3 Flow control device 4 Level sensor Fig. 3: Safety equipment Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 12: Guards

    Gesundheitsschutzkennzeichnung am Arbeitsplatz". Caution labels on the unit must be easily readable at all times. Illegible caution labels must be exchanged without delay. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 13: In Case Of Accidents

    Should you or another person be injured when working with the unit: • Stay calm! • Render first aid! • Call the company first-aider without exception! Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 14: Product Description

    4.1 Intended Use The liquid-air cooler LA 5000 is used for the cooling of a liquid circuit. As a coolant water or a water/glycol mixture may be used. The coolant circulates between the cooling unit and the device to be cooled. It is re- cooled by an air-cooled heat exchanger.
  • Page 15: Unit Components

    The liquid temperature is controlled by an electric thermostat, whereas liquid throughput is controlled by a flow control device. Both indications are made potential-free and must be integrated into the safety circuit of the device to be cooled. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 16: Specifications

    Current draw: 3.3 A Customer main fuse: 5 A T Safety circuit, max input 230VAC, 0,5A, 15W 100VDC, 0,5A, 10W Table 8: Electrical characteristics Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 17: Setting-Up Requirements

    The following infrastructure is required for connecting the unit: Parameter Rated value Operating voltage 230±10% VAC, 1Ph, 50/60 Hz Table 9: Required infrastructure Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 18: Transport

    Lift the unit with a forklift or jack lift off the transportable pallet. Dispose of the packaging material in accordance with regional regulations. PLEASE NOTE LAIRD THERMAL SYSTEMS advises to keep the transportable pallet for later transportation of the unit. Laird Thermal Systems s.r.o Date: 08/09/2022...
  • Page 19: Initial Operation

    In case of storage of the unit at temperatures lower than 5°C or higher than 40°C for longer periods please wait 3 hours prior to initial operation to allow for temperature adjustment. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 20: Cooling Circuit Connection And Filling

    Transparent hoses stimulate algae growth that may increase the error-proneness of the components built into the unit. Thus only use non-transparent hoses. Fig. 9: Cap of coolant container 1 Cap Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 21: Electrical Connections

    Required information can be drawn from the specifications listed on page 16 and the wiring diagram in the addendum. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 22: Carrying Out Setting To Work

    After connecting the cooling circuit, filling the coolant container and finishing the electrical connections follow the steps below for setting the unit to work: Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 23: Daily Start-Up

    6.4 Setting to work after Storage Setting to work after storage will have to follow the same procedures as required for initial operation (see chapter 6.2). Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 24: Controlling The Unit

    Close all valves that may exist in the extension of the hoses running to and from the unit.  The unit is out of operation. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 25: Settings

    By-pass valve The by-pass valve is set by the manufacturer to a maximum pressure of 9 bars. If any modification to this setting should be required, please contact the LAIRD THERMAL SYSTEMS service department to receive briefing. Before adjusting the flow control device, the fan’s power must be disconnected! Setting the Thermostat Laird Thermal Systems s.r.o...
  • Page 26  The switch-off temperature is set to a higher value. Decrease the temperature setpoint Turn the knob counterclockwise.  The switch-off temperature is set to a lower value. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 27: Disruptions

    More help can be found in the following paragraph. In case you do not succeed in identifying the problem cause by means of this manual please contact the service department of LAIRD THERMAL SYSTEMS. 8.1.1 Trouble Shooting For trouble shooting you may rely on the following: •...
  • Page 28: Maintenance And Cleaning

    Since the cooling circuit is an open circuit, evaporation of coolant may occur. Thus the filling level of the coolant container has to be checked regularly and coolant might have to be refilled as described on page 23. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 29: Cleaning Of Strainer

    Risk of damage to the pump, if the unit is operated with a fully closed ball valve! Never start the unit, when the ball valve is closed. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 30: Cleaning Of Unit Casing

    Regularly remove dirt from the casing of the unit to prevent corrosion damage and clogging of the air grids. Pay attention that all the plates at the unit are always clean and legible. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 31: Repair

    10 Repair 10.1 Repair of unit In case of malfunction during the warranty period the unit must be sent to the LAIRD THERMAL SYSTEMS service department for repair (see page 7). When warranty has expired, no restrictions from the side of LAIRD THERMAL SYSTEMS exist with respect to repair work carried out by the customer as long as guarantee and warranty conditions remain untouched.
  • Page 32: Dismounting, Disposal, Storage

    The coolant has to be collected and disposed of according to valid regulations. Let the coolant container run empty into an appropriate vessel. Clean the unit.  The unit is placed out of operation. Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 33: Drain Of Unit

    Make sure the components of the unit end up at a qualified company for disposal and recycling. Contact LAIRD THERMAL SYSTEMS for take back of end-of-life units (see company contact on page 7 or ask an qualified company for disposal and recycling.
  • Page 34: Return Of The Unit To Laird Thermal Systems

    Return of the unit to LAIRD THERMAL SYSTEMS 11.7 Return of the unit to LAIRD THERMAL SYSTEMS PLEASE NOTE Declaration of decontamination Before re-shipment of the unit a declaration of decontamination must be sent to LAIRD THERMAL SYSTEMS. Laird Thermal Systems s.r.o Date: 08/09/2022...
  • Page 35: Wear Parts And Spare Parts

    LAIRD THERMAL SYSTEMS parts are subject to strict obligations and fulfill these requirments. LAIRD THERMAL SYSTEMS does not provide warranty service in case of damages caused by the use of spare parts made by manufacturers other than LAIRD THERMAL SYSTEMS.
  • Page 36 387002780 387005416 Motor 95205201.00 Flexible coupling 95205203.00 387007783 Level switch 95240301.00 Strainer (inside pump casing) 96299012.00 Flow switch 95140538.01 Table 12: Spare parts Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 37: Addendum

    Pump Relief valve, part of pump Strianer, part of pump Tank for coolant Heat exchanger Ball valve Level switch Flow switch Thermal switch Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 38: Wiring Diagram

    Addendum Wiring diagram Wiring diagram Legend Level switch Flow switch Thermal switch Pump Resistor Resistor Terminal block Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 39: Dimensional Drawing

    Addendum Dimensional drawing Dimensional drawing Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 40: Performance Diagram

    Addendum Performance diagram Performance diagram Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...
  • Page 41: Flow Versus Pressure

    Addendum Flow versus pressure Flow versus pressure Laird Thermal Systems s.r.o Date: 08/09/2022 Liquid-Air Cooling Unit LA 5000 Operation Manual, 387009851 387009381 Version: 1.0...