Wiring; Power Supply; Signal Output - Intec AIRSENSE I-310e Operator's Manual

Microprocessor-based, infrared environmental co2 sensor
Table of Contents

Advertisement

I-310e – UserManual

Wiring

This section describes the external connections to the Model
I-310e. Wiring enters the chassis through the cutout in the
center of the unit.

Power Supply

The Model I-310e will operate from an AC or DC input
voltage between the values called out in the specifications on
page 2. The power supply leads are connected to the two-
terminal power connector shown in Figure 2.
The model I-310e must never be connected directly to
line power. Operation at voltages higher than specified
will damage the unit and void the warranty.
When operating from DC power, the polarity of the power
leads must be as shown in Figure 2. Reversed polarity
connection will not damage the unit, but will make it
inoperable until the connection is reversed.

Signal Output

The Model I-310e provides either a 0 - 10 volt or a 4 - 20
mA current loop output at the two terminals of the analog
output connector shown in Figure 2.
INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com
Specifications subject to change without notice. |
J3
USER
JUMPERS
JP5
JP4
JP3
UP
JP2
BUTTON
JP1
UP
DIS1
SHORTING BLOCK HERE
FOR CURRENT OUTPUT
MICROPROCESSOR
ANALOG OUTPUT
SELECTOR
SHORTING BLOCK HERE
DN
FOR VOLTAGE OUTPUT
J4
-
DOWN
+
BUTTON
J8
ANALOG OUTPUT
RELAY CONNECTION
Figure 2: Model I-310e Component Locations
DCS, Inc. | PRE 200217 | Page 7 of 17
©
UPPER MOUNTING
SCREW ACCESS
WIRING
ACCESS
POWER IN
(POLARITY MATTERS
+
FOR DC INPUT ONLY)
J5
LOWER MOUNTING
SCREW ACCESS