Duct Kit Contents

1
1

Duct Kit Installation

1) Select a point along the duct where the probe assembly can
be installed into unrestricted airflow without interfering with
any internal duct components such as dampers, radiators, etc.
2) Mark and drill the four holes for the duct probe as shown in
Figure 4. The centerline must be parallel to the air flow
through the duct.
3) Install the duct probe assembly through the holes just drilled.
The sample port must be on the upstream side. Secure
the probe assembly in place with the two sheet metal screws.
4) Connect the open ends of the two tubes from the probe
assembly to the two sample ports on the base of the Model
310e. It makes no difference which tube is connected to
which port on the Model 310.
INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com
Specifications subject to change without notice. |
dust filter
duct probe assembly
Figure 4: Duct Probe Installation Dimensions