• Corrosive substances, such as halogenated hydrocarbons. • Catalytic poisons, such as sulfur and phosphor compounds, silicon compounds, metal vapors. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
• Corrosive substances, such as halogenated hydrocarbons • Catalytic poisons, such as sulphur and phosphor compounds, silicon compounds, metal vapours INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
30-90% of the measuring range. Rest is synthetic air. • Extraction set consisting of gas pressure regulator and flow meter • Calibration adapter with tube; CONKIT-PG2. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
• Symptoms: Dizziness, headache and nausea. • Procedure if exposed: Remove victim to fresh air at once, seek medical attention. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Calibration will not be possible when the sensitivity of the sensor drops to 40% or lower; the Sensor Cartridge must be replaced. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Remove dust deposits, especially at the gas inlet. • The filter at the gas inlet has to be replaced if extremely dirty. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
(adjustable in system parameters -> error time, value range> 1) until the calibration is repeated. Performing the calibration automatically deletes this message. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Please check the SC2 datasheet for the latest specifications. Available at: http://inteccontrols.com/gas-detection/SC2.html http://inteccontrols.com/gas-detection/SC2.pdf 9.2 AT6 Please check the AT6 datasheet for the latest specifications. Available at: http://inteccontrols.com/gas-detection/AT6.html http://inteccontrols.com/gas-detection/AT6.pdf INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Page 14
The sensitivity of Pellistor sensors can be influenced by substances containing silicon compounds and even poisoned/destroyed by them. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
1 The table doesn’t claim to be complete. Other gases, too, can have an influence on the sensitivity. The mentioned cross sensitivity data are only reference values valid for new sensors. 2 Cross sensitivities valid for all measuring ranges of the sensor. INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
13.3 Limited Warranty INTEC Controls warrants the SC2 / AT6 devices for a period of one (1) year from the date of shipment against defects in material or workmanship. Should any evidence of defects in material or workmanship occur during the warranty period, INTEC Controls will repair or replace the product at their own discretion, without charge.