INITIAL STARTING PROCEDURE

Until you familiarize yourself with the use of all controls,
the following procedure may be used to standarize the in-
itial setting of controls as a reference point and to obtain
CS-1022, 1 0 2 1 , 1 0 1 2

( 1 ) NORMAL SWEEP DISPLAY OPERATION

1. Turn the POWER control
supply will be turned on and the pilot lamp will light.
Set these modes as follows;
MODE ©
CH1
:
TRIG MODE © : AUTO
2. The trace will appear in the center of the CRT display
and can be adjusted by the CH1 • POSITION ©
o
POSITION
®
controls. Next, adjust the INTENSI-
TY
©
and, if necessary, the FOCUS ©
observation.
3. Vertical Modes
W i t h vertical MODE
signal t o the CH1
INPUT
VOLTS/DIV (2) control for a suitable size display of the
waveform. If the waveform does not appear in the
display, adjust the VOLTS/DIV and y POSITION controls
to bring the waveform into the center portion of the CRT
display. Operation w i t h a signal applied to the CH2 IN-
PUT
(D
jack and the vertical MODE set to CH2 is
similar to the above procedure.
In the ADD mode, the algebraic sum of CH1 + C H 2 is
displayed. If the CH2 INV ©
ed, the algebraic difference of the t w o waveforms,
CH1 — CH2 is displayed. If both channels are set to the
same VOLTS/DIV, the sum or difference can be read
directly in VOLTS/DIV from the CRT. In the ALT mode,

OPERATION

©
clockwise — the power
for ease of
©
set to C H 1 , apply an input
(5) jack and adjust the
switch has been engag-
trace on the CRT in preparation for waveform observation.
When using the probe(s), refer to probe's instructions and
"PROBE COMPENSATION" listed in APPLICATION of this
manual.
Fig. 7
( 1 ) NORMAL SWEEP DISPLAY OPERATION
1. Turn the POWER control
supply will be turned on and the pilot lamp will light. Set
the TRIG MODE ©
2. The trace will appear in the center of the CRT display
and can be adjusted by the y POSITION 0
POSITION
©
and
®
and, if necessary, the FOCUS
observation.
3. Apply an input signal to INPUT © jack and adjust the
VOLTS/DIV (2) control for a suitable size display of the
waveform. If the waveform does not appear in the
display, adjust the VOLTS/DIV and v POSITION controls
to bring the waveform into the center portion of the CRT
display.
4. If no trace is obtainable, refer to the following TRIGGER-
ING procedures.
The display on the screen will probably be unsynchroniz-
ed.
Refer to TRIGGERING procedures below adjusting syn-
chronization and sweep speed to obtain a stable display
showing the desired number of waveforms.
TRIGGERING
The input signal must be properly triggered for stable
waveform observation. TRIGGERING is possible the input
signal INTernally to create a trigger or with an EXTernally
CS-1020, 1 0 1 0
©
clockwise—the power
to AUTO position.
controls. Next, adjust the INTENSITY
©
for ease of
and
15