OPERATION PROCESS

(1) Turn on the main switch.
Product remains as standby mode having memory in Program 1 like below picture when power on.
(2) Choose the program.
Press the bottom of the left side button
on the foot controller or
button on the control box,
the program changes by 1 stage. Select the program that user want to use.
If press the
button on the foot controller when the program is selected program 9,
it returns to program 0.
(3) Check the setting.
Check the display on LCD box to make sure the setting is correct.
(4) Start the Operation.
The motor starts to rotate after press the speed control pedal. It runs at low speed with light press and
at high speed with heavy press by the set speed in max. If the irrigation mode is on, the irrigation pump
operates too.
(5) Overload protection function.
If the load exceeds the set torque during the operation, the overload protection function works not to
generate the extra torque.
(6) Stop the operation.
The operation is stopped when user takes foot off the speed control pedal.
(7) Change the operation direction.
Press the
button on the foot controller or
button on the control box to reverse direction.
In reverse direction mode, the device will sound beep sound, and it continues during the operation.
(If one more press the button, it is changed to forward direction mode.)