(3) ELECTROMAGNETIC IMMUNITY / RADIOFREQUENCY MOBILE EQUIPMENT
Conducted RF
rms
IEC 61000-4-6
EN 61000-4-6
Radiated RF
IEC 61000-4-3
EN 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strength of fixed radiofrequency emitters, such as base stations
for mobile telephones (cellular / cordless), mobile radio, AM/FM radio broadcasts and TV
broadcasts cannot be determined exactly by theory.
To assess the electromagnetic environment due to fixed radiofrequency emitters, an
electromagnetic environment measurement must be made. If the measured radiofrequency field
strength in the immediate environment where the product is used exceeds the compliance level
specified above, the performance of the product must be tested to verify whether it conforms
to the specification. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be
less than 3V/m.
Micromotor with Motor cord
Foot controller with cord
Portable and mobile RF communications equipment should be used
no closer to any part of the
rms
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80MHz to 800MHz
d = 2.3 √P 800MHz to 2.5GHz
Where 'P' is the maximum output power rating of the transmitter in watt
(W) according to the transmitter manufacturer, and 'd' is the recommended
separation distance (m).
The electromagnetic field strength of fixed radiofrequency emitters, which is
determined by an electromagnetic environment measurement (a), must be
less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Maximum length
1.2M
1.2M
1.5M
X-CUBE
including cables, than the
: IEC 61000-3-3 / EN 61000-3-3
: IEC 61000-4-8 / EN 61000-4-8