Program Setting
X-CUBE has 9 programmable memories in conditions below
Set the gear ratio – speed – torque – operation direction – irrigation pump. And press the memory button.
ex)
Error and Remedy
If the motor is stopped due to the overload, user's error and etc., circuit safety system cuts the power
supply and shows an error code on LCD window.
Indication
Warning!
Torque LIMIT.
Warning!
Motor Connect Check!
Warning!
Foot Switch Check!
Cause & Remedy
· In case that the actual torque reaches preset torque and
keeps it more than 2 seconds.
- Remedy: Release the foot controller.
· In case that the motor is not connected properly.
- Remedy: Make sure to connect Motor properly.
Press the foot controller.
If the remedy does not work, contact the manufacturer.
· In case that the foot controller is not connected properly.
- Remedy: Make sure to connect the foot controller.
Press the foot controller.
If the remedy does not work, contact the manufacturer.