2. Sterilization
Sterilization part list
· Angle handpiece
· Motor (assembling with the motor cap)
Sterilization method
① Dry the applicable products completely after cleaning.
② Put the applicable products in a sterile bag and seal tightly.
※ Make sure to assemble the motor cap for autoclave on the motor.
※ Do not disassemble the motor and motor cord when sterilizing.
③ Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132℃, and let it dry for
30 minutes.
Sterilization
Moist heat sterilization
* Maximum sterilization number is 250 times.
* Make sure to use the sterilization bag and seal tightly.
* Make sure to dry the applicable products before sterilization.
* Motor could be damaged in sterilization because of the moisture inflow or else, so should be
careful to handle it.
* Sterilizing the angle handpiece with blood or other remains may cause the breakage of the angle
handpiece. Therefore, it is necessary to clean the angle handpiece before sterilization.
* Irrigation tube and Y-connector are disposable. Do not use them again.
3. Lubrication
Lubrication product list
· Angle handpiece
Lubrication method
- Refer the angle handpiece manual.
* Do not insert the oil into the motor. It damages the bearing or causes fever.

ATTENTION

1. Please read the notice below before operating X-CUBE.
1) Check the connection of every cords and correct operation when power-on.
2) Check the earth connection is safe.
3) Make sure to check the contact part to patient.
4) Check the voltage.
5) Check the sterile condition of autoclavable parts.
2. Please read the notice below during X-CUBE use.
1) Check the status of product and patient constantly during the surgery.
2) When product or patient has any problem, stop the operation in condition of patient's safety or take
proper action.
3) Do not let patient to approach the product.
3. Please read the notice below after operating X-CUBE.
1) Do not put any physical pressure on the power cord, motor cord and foot controller cord.
Temperature
132℃
Time
4 minutes
132
Dry
30 minutes
34