Printer Function Messages

If tympanometric results are not printed, a message will appear describing
the reason. These messages are listed in Table 2.
Printer Function
Message
No Data
Reinsert Handle
No Data
Nothing in Memory
Computer Interface Switch #4 on
Switch 4 is Set on
Bottom of Printer
Table 2 – Printer Function Messages
Possible Cause
The MicroTymp Handle
is not located in the well. the well.
The MicroTymp Handle
is not seated properly
in the well.
The MicroTymp Handle
has a discharged or
missing battery.
The MicroTymp Handle
is not functioning
properly.
Both right
left
memory loca-
L
tions in the MicroTymp
Handle are empty.
Printer/Charger is ON.
Possible Solution
Seat the handle in
Insure that the handle
is fully seated in the well
with the Liquid Crystal
Display (LCD) and
buttons facing you.
Verify battery is in
place, and charged
(LOW BATT message
does not appear).
Call your nearest
Welch Allyn service
location, distributor,
or factory representative.
and
Insure that data is
R
being stored correctly.
See page 7.
Turn switch #4 OFF.
See page 22.