7 Reprocessing steps in accordance with ISO 17664
Note
Please comply with the instructions for use of the washer disin-
fector.
▶ In order to prevent impairment of the KaVo medical device,
make sure that the inside and outside of the device is dry
after the end of the cycle.
▶ Remove any residual liquids with KaVo DRYspray.
See also:
2 7.2.5 Manual drying, Page 63
▶ Immediately after drying, lubricate the KaVo medical device
with care agents from the KaVo care system.