7 Reprocessing steps in accordance with ISO 17664
55 / 268

7.2 Manual Reprocessing

WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶ Remove the tool.
NOTICE
Never reprocess this medical device in an ultrasonic
cleaner.
Malfunction and material damage.
▶ Clean manually or in a washer disinfector only.