Option
LEVEL UNITS
PRIMARY DEVICE
PROGRAM LOCK
SAMPLER PACING Enables or disables sampling. Sets the sample interval. Options: 100 gallons (gal),
SITE ID
TOTAL FLOW
UNITS
VELOCITY
DIRECTION
VELOCITY UNITS
VELOCITY
CUTOFF
VELOCITY
DEFAULT
Option
Description
gps
Gallons per second
gpm
Gallons per minute
gph
Gallons per hour
lps
Liters per second
lpm
Liters per minute
lph
Liters per hour
Configure data logging
Select the input channels that are recorded to the data log.
Note: No readings are recorded to the data log until data logging is configured.
1. Push MAIN MENU.
2. Select OPTIONS>ADVANCED OPTIONS>DATA LOG.
Description
Sets the measurement units for level.
Selects the primary device.
Enables or disables the program lock. The program lock prevents unauthorized use of
the keyboard and access via RS232 or modem. The program lock password is
9500 and cannot be changed.
liters (ltr), cubic meters (m
Sets the site ID (maximum of 8 digits). The site ID is on all data printouts. Use this
feature when multiple sites are monitored with a single flow meter or if data readings
from multiple flow meters are collected.
Note: A text site ID can be set with InSight data management software and an
RS232 connection.
Sets the measurement units for total flow. Options: gallons (gal), liters (ltr), cubic
3
meters (m
), acre-feet (af) or cubic feet (cf)
Sets the direction of velocity. Options: UPSTREAM (NORMAL), DOWNSTREAM or
ALWAYS POSITIVE
Sets the measurement units for velocity. Options: ft/s, m/s
Sets the velocity cutoff. Use when the site has low velocities and frequent low
particulate concentrations that prevent velocity measurements.
Example 1: Velocity cutoff = 0.20 ft/s, Velocity default = 0 ft/s
If the velocity is less than 0.20 ft/s, the meter saves a value of 0 ft/s until the velocity
increases to more than 0.20 ft/s.
Example 2: Velocity cutoff = 0.20 ft/s, Velocity default = 0.20 ft/s
If the velocity is less than 0.20 ft/s, the meter saves a value of 0.20 ft/s until the velocity
increases to more than 0.20 ft/s.
Sets the velocity value that is used when velocity cannot be measured.
Table 15 Flow unit options
Option
Description
mgd
Million gallons per day
afd
Acre-feet per day
cfs
Cubic feet per second
cfm
Cubic feet per minute
cfh
Cubic feet per hour
3
), acre-feet (af) or cubic feet (cf)
Option
cfd
cms
cmm
cmh
cmd
Description
Cubic feet per day
Cubic meters per second
Cubic meters per minute
Cubic meters per hour
Cubic meters per day
English 21