Picture 5
Attach the luer-lock (N) to the pump by tightening it into the pump adapter.
Make sure there is a cartridge containing sufficient insulin in the cartridge
compartment. Remove the circular protective cap (L) by pressing the side clips
inwards gently. Follow the pump manufacturer's user manual to prime the
transfer-set until medication drips from the connector needle tip. Do not leave
air bubbles in either the cartridge or the transfer-set.
Picture 6
Connect the transfer-set to the newly inserted head-set with the bulging side
upwards. A click must be heard when the connection engages. Fill the empty
space of the cannula housing (H) with a bolus of 0.7 units of insulin U-100
(about 7μl). Start distribution with the pump.