6. Dry device with a lint-free single-use cloth.
7. Pay special attention to device seams and insertions.
4.2.2 Manual Disinfection
1. Thoroughly wipe cleaned device [4.2.1 Manual Cleaning] using a single use cloth
dampened with a bactericidal, virucidal, fungicidal, alcohol-based disinfectant so-
lution approved according to local regulations and use according to disinfectant
manufacturer's Instructions for Use. Please note [4.2] for wrong cleaning or disinfec-
tion method.
2. Pay special attention to device seams and insertions.
3. Remove disinfectant solution residue with a damp cloth.
4. Dry device with a lint-free single-use cloth.
5. Pay special attention to device seams and insertions.
Device can be disinfected for up to 6,000 times.
4.3 SmartLite
 FOCUS
®
NOTICE
Wrong cleaning or disinfection method.
Damage to SmartLite
1. Do not clean or disinfect with highly aggressive agents (e.g. solutions based on
orange oil or acetone).
4.3.1 Manual Cleaning
1. Discard examination gloves according to local regulations and disinfect hands with
an appropriate bactericidal, virucidal, and fungicidal hand disinfectant solution ac-
cording to local regulations and use according to disinfectant solution manufac-
turer's Instructions for Use.
2. Use a new clean pair of examination gloves.
3. Clean device with an appropriate cleaning solution according to local regulations
and use according to cleaning solution manufacturer's Instructions for Use.
4. Pay special attention to device seams and insertions.
5. Remove cleaning solution residue with a cloth soaked with water.
6. Dry device with a lint-free single-use cloth.
4.3.2 Manual Disinfection
1. Thoroughly wipe cleaned device [4.3.1 Manual Cleaning] using a single use cloth
soaked with a bactericidal, virucidal, fungicidal, alcohol-based disinfectant solution
approved according to local regulations and use according to disinfectant manufac-
turer's Instructions for Use.
2. Pay special attention to device seams and insertions.
3. Remove disinfectant solution residue with a cloth soaked with water.
4. Dry device with a lint-free single-use cloth.
SmartLite
 FOCUS
®
4.3.3 Sterilization (optional)
The SmartLite
®
1. Steam autoclave SmartLite
3 minutes 30 seconds.
SmartLite
 FOCUS
®
Shields
®
6
 FOCUS
®
®
Shields can be disinfected for up to 100 times.
®
 FOCUS
Shields can also be autoclaved.
®
 FOCUS
®
Shields can be autoclaved for up to 60 times.
®
Shields.
Shields at 134 °C/2 bar = 273 °F/29 psi for
®
8