Abr screening

Advertisement

Bio-logic
NavPRO ONE
®
Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making
sure that the ear tip fits tightly into the ear canal.
Select QuickABR button
Select the test ear by pressing either
the Right or Left button
A calibration will be performed prior
to measurement.
A green light indicates
successful calibration.
An impedance check
will be performed prior to
measurement. An impedance
below 6 k ohm is recommended
for best test results.
®
4
QUICK START GUIDE
After successful ear probe calibration
and impedance check, the test will
start automatically. If yellow or
red lights are displayed, check to
make sure you are using the correct
probe tip and ear tip size, that the
eartip is inserted correctly, and that
there is no excessive ambient noise.