Product Description - Tektronix TDS 500B Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 500B:
Table of Contents

Advertisement

Specifications

Product Description

TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
This section begins with a general description of the traits of the TDS 500B,
TDS 600B, and TDS 700A oscilloscopes. Three sections follow, one for each of
three classes of traits: nominal traits, warranted characteristics, and typical
characteristics.
The TDS 500B, TDS 600B and TDS 700A Digitizing Oscilloscopes are
portable, four-channel instruments suitable for use in a variety of test and
measurement applications and systems. Table 2–1 lists key features.
Table 2–1: Key Features of the TDS 500B, 600B and 700A Oscilloscopes
Feature
TDS 600B
Digitizing rate,
TDS 684B: 5 GS/s on ea. of 4 ch
maximum
TDS 680B: 5 GS/s on ea. of 2 ch
TDS 644B: 2.5 GS/s on ea. of 4 ch
TDS 620B: 2.5 GS/s on ea. of 2 ch
Analog bandwidth
TDS 680B, 684B, 782A and 784A: 1 GHz
TDS 520B, 540B, 620B, 644B, 724A and 754A: 500 MHz
No. of Channels
TDS 644B & 684B: 4
TDS 620B & 680B: 2 + 2
Record lengths,
15,000 samples
maximum
Acquisition modes
Sample, envelope, peak detect
and average
Trigger modes
Include: edge, logic, and pulse.
Video trigger, with option 05, modes include:
NTSC, SECAM, PAL, HDTV, and FlexFormat.
Display
TDS 520B, 540B, 620B, 680B: Monochrome
TDS 644B, 684B, 724A, 754A, 782A, 784A: Color
Storage
1.44 Mbyte, 3.5 inch, DOS 3.3-or-later floppy disk (optional on
TDS 520B, 540B, 620B & 680B).
NVRAM storage for saving waveforms, hardcopies, and setups
I/O
Full GPIB programmability.
Hardcopy output using GPIB, RS-232, or Centronics ports
1
Two plus Two channel operation allows up to two of the four channels to be
displayed simultaneously. Channels not displayed can be used to couple a
triggering signal to the oscilloscope.
TDS 500B & TDS 700A
TDS 784A: 4 GS/s
TDS 540B, 754A, 782A: 2 GS/s
TDS 520B, 724A: 1 GS/s
Opt. 1G, TDS 540B: 1 GS/s
TDS 540B, 754A & 784A: 4
1
TDS 520B, 724A & 782A: 2 + 2
50,000 samples
(500,000 with option 1M)
Sample, envelope, average,
high-resolution, and
peak-detect
1
2–1