Amplitude Accuracy Across Frequency Verification (Continued - Anritsu Spectrum Master MS2721B Maintenance Manual

Handheld spectrum analyzer and base station analyzer
Hide thumbs Also See for Spectrum Master MS2721B:
Table of Contents

Advertisement

MS2721B
Firmware Revision: ________________ Operator: ________________ Date: ______________
Serial Number: ________________ Options: ___________________________________________________

Amplitude Accuracy Across Frequency Verification (continued)

Table A-29. Amplitude Accuracy Across Frequency Verification at 4000 MHz
Input Power
Attenuation Level
–30 dBm
–30 dBm
–30 dBm
–30 dBm
–2 dBm
–2 dBm
–2 dBm
–2 dBm
Table A-30. Amplitude Accuracy Across Frequency Verification at 5000 MHz
Input Power
Attenuation Level
–30 dBm
–30 dBm
–30 dBm
–30 dBm
–2 dBm
–2 dBm
–2 dBm
–2 dBm
Table A-31. Amplitude Accuracy Across Frequency Verification at 6000 MHz
Input Power
Attenuation Level
–30 dBm
–30 dBm
–30 dBm
–30 dBm
–2 dBm
–2 dBm
–2 dBm
–2 dBm
MS2721B MM
0 dB
5 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
0 dB
5 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
0 dB
5 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
PN: 10580-00177 Rev. D
Marker 1 Reading
Marker 1 Reading
Marker 1 Reading
Specification
dBm
± 1.25 dB
dBm
± 1.25 dB
dBm
± 1.25 dB
dBm
± 1.25 dB
dBm
± 1.25 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
Specification
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
Specification
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
dBm
± 1.75 dB
A-11