Physical Characteristics
Weight (probe only)
Cable length
Calibration
This section contains calibration information and verification procedures for the
SA1000X.
Verification confirms whether the probe is operating within published
specifications. If the probe performs outside of the limits defined in the verification
procedures, return the probe to NI for repair.
Refer to
NI Services
Related tasks
When to Calibrate the SA1000X
■
Related reference
Electrical Specifications
■
Test Equipment
Refer to
Table 1
model recommendations for calibration of the SA1000X.
If you do not have the recommended equipment, select a substitute calibration
standard using the specifications listed in the Requirements column.
Equipment
2 × DMM
12
ni.com
for more information on contacting NI.
and
Table 2
for a list of necessary equipment and, where applicable,
Recommended
Model
NI PXIe-4082
SA1000X Safety, Environmental, and Regulatory Information
Where Used
Requirements
Verifying DC
Accuracy
(× 2)
96 g
1.3 m
Measurement error:
■
<125 ppm
Resolution: <1 µV
■