...7 PROGRAMMING

7.1 Access to Secure Parameters

0 0 0 0 0
SECURITY CODE
– – – – –
Espanol

7.2 Language Page

_ _ _ _ _
Espanol
˜
Francais
Deutsch
English
– – – – –
SET UP PARAMETER

7.3 Set Up Parameters Page

– – – – –
SET UP PARAMETER
– – – – –
Type pH Antimony
pH Glass
Redox (ORP)
Redox
pH
24
Security Code
Enter the required code number, between 00000 and 19999, to gain access to the
secure parameters. If an incorrect value is entered, access to subsequent programming
pages is prevented and the display reverts to the
Advance to
Language Page
Language Page
Select the language to be displayed on all subsequent pages.
Advance to
Set Up Parameters Page
– press to advance to next parameter
or
– press to advance to next page.
These two keys are used to advance to all subsequent parameters and pages. If a
parameter is changed it is automatically stored on operation of either key.
Probe Type
Select pH Antimony/pH Glass for pH display or Redox (ORP) for millivolt display.
Select the required probe type.
Continued on next page.