Adjust RTX Zero
– – – – –
Adjust RTX Span
– – – – –
Adjust RTX Zero2
– – – – –
Adjust RTX Span2
Return to
top of page
0 0 0 0 0
Alter Fact. Code
7 0 0
Monitoring pH
2 5 0
Redox (ORP)
32
Adjust Retransmission Zero
Set the milliammeter reading to 4mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Adjust Retransmission Span
Set the milliammeter reading to 20mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Zero (available only on 4630/800, 4631/800, 4635/800 and
4636/800 instruments)
Set the milliammeter reading to 4mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Span (available only on 4630/800, and 4631/800, 4635/800
and 4636/800 instruments)
Set the milliammeter reading to 20mA.
Note. The retransmission range selected in the
not affect the reading.
Alter Factory Settings Access Code
Set the factory settings access code to a value between 00000 and 19999.