Control Panel

Thermo Scientific

Features

Section 5
Power
Hi-Limit
On
Off
1
2
Figure 5-1. Control Panel
1. Power Switch: Turns all power to the unit on or off.
2. Power Switch Status Lamp: When power is on, this green lamp is lit.
3. Hi-Limit Thermostat: This hydraulic type thermostat provides
mechanical overtemperature back-up control if the temperature
programmer-controller should fail.
4. Hi-Limit Thermostat Status Lamp: When lit, this red lamp signals that
the temperature is exceeding the hi-limit setpoint established by the
user.
5. Programmer-Controller: 4-button operation controller allows user to
set and control operating temperature and program ramp-to-setpoint
procedures (operating details follow in the upcoming OPERATION
section).
3
4
AL1
AL2
REM MAN
A-T
SP
RAMP
OP
DWELL
TIME
HOLD
RUN
REM
P
HOLD
5
Clean Room Oven
5-1

Advertisement