Latch Adjustment -Heating Door Opens - Thermo Scientific 3490M Manual

Table of Contents

Advertisement

Latch Adjustment -
Heating Door Opens
Thermo Scientific
If the door opens while oven is heating, follow the procedure below to
adjust the latch. The door will have to be adjusted to close more easily to
allow for the expansion occurring during heating.
1. Open the door. Use 5/64" Allen wrench to remove the Allen setscrew
from the door strike assembly.
2. Rotate catch mechanism counterclockwise to increase length and,
thereby, position door a little farther away from the body of the oven.
The number of turns will have to be estimated. Re-insert and tighten
Allen setscrew securely.
3. Repeat if necessary to achieve satisfactory operation.
Depending on operating temperature, the procedure may have to be
repeated to ensure proper door operation.
Section 4
Unpacking and Installation
Clean Room Oven
4-5