...6 PROGRAMMING

6.2.5 Current Output Page

The current output is assigned to the CO
Fig. 2.4.
CURRENT OUTPUT
OP1 Cal Hold
Enter
OP1 Law
Enter
– – –
OP1 FSD
or
Enter
Lin
Log
OP1 Zero
or
Enter
18
concentration but is only operative if the relevant output modules are fitted – see
2
Page header.
Advance to next parameter.
Yes
Output 1 Calibration Hold
Current Output 1 can be held during calibration, if required. Select 'YES' or
or
No
'NO'.
Store.
Advance to next parameter.
Log
Output 1 Law
Current Output 1 can be either logarithmic or linear. Select 'Log' or 'Lin'.
or
Lin
Store.
Advance to next parameter.
Output 1 Full Scale
Current Output 1 full scale range: 1 to 1000mgl
Set the required concentration value for Current Output 1 full scale.
Store.
Advance to next parameter.
– –
Output 1 Zero
Set the required concentration value for Current Output 1 zero. Default is
FSD.
Note. If the output 1 Law is linear (Lin), this parameter is omitted and the zero
current output is automatically set to '0'.
Store.
Advance to next parameter.
Advance to next parameter.
Continued on next page.
–1
CO
, i.e. minimum span 1 decade.
2
1
/
of
10