Translation of the original operating manual Freeze-dryer Rev. 3.0 of 20/07/2022 Epsilon 1-4 LSCplus;Freeze-dryer;Epsilon 2-4 LSCplus; ; Version 07/2022 Epsilon 1-4 LSCplus Part no. 111141 Epsilon 2-4 LSCplus Part no. 111144 Freeze-dryer Operating Manual Read the manual thoroughly before use and retain it for later use!
Page 2
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Version 07/2022, Rev. 3.0 of 20/07/2022 • sb 2 / 152 Translation of the original operating manual...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Table of contents General information ......................9 Importance of the operating manual .................. 9 Further applicable documents ................... 9 Intended use ........................9 1.3.1 Freeze-drying of solvent-containing products (non-aqueous media) ......11 1.3.2...
Page 6
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Table of contents Safety notes concerning the operation ................30 3.7.1 Hazards caused by electricity ..................30 3.7.2 Hazards caused by natural, flammable refrigerants ........... 30 3.7.3 Hazards caused by nitrogen ..................31 3.7.4...
Page 7
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Table of contents Operation ..........................53 Initial start-up ........................53 Installation of accessories ....................53 Preparation ........................53 Switching the freeze-dryer on ..................53 Loading door ........................54 LSCplus control system ....................56 6.6.1...
Page 8
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Table of contents Maintenance and service ....................114 Maintenance ......................... 114 8.1.1 General ........................114 8.1.2 Chamber ........................115 8.1.2.1 Lamination of the loading door with a special film ..........116 8.1.2.2 Door latch ......................
Martin Christ Gefriertrocknungsanlagen GmbH have been performed. • The vacuum pump used is approved by Martin Christ Gefriertrocknungsanlagen GmbH. The vacuum pumps approved for this system are listed in (see chapter 5.2.3 - "Vacuum pumps"). •...
Page 10
Exclusion of liability Any use of the freeze-dryer other than the one described here or any use going beyond this description is regarded as improper use. Martin Christ Gefriertrocknungsanlagen GmbH cannot be held liable for any damage resulting from such improper use.
Freeze-drying of azide-containing products Freeze-drying of products containing azides is only permissible following a case-by-case examination and the written approval by Martin Christ Gefriertrocknungsanlagen GmbH, since azides may form explosive dust- air-mixtures and also explosive metal azides when combined with non- ferrous metals (e.g.
Pos : 16 /200 C hrist/360 GT-BA Labor-Pilot (ST ANDARDM ODU LE)/010 Allgem eine Informationen/010- 0040 Ur heberrec ht @ 25\mod_1404983212403_68.doc x @ 182977 @ 2 @ 1 → Copyright The copyright concerning the operating manual remains with Martin Christ Gefriertrocknungsanlagen GmbH. The operating manual is solely intended for the operator and their personnel.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation Pos : 25 /200 C hrist/360 GT-BA Labor-Pilot (ST ANDARDM ODU LE)/020 Aufbau und Wirk ungsw eise/020 Aufbau und Wirk ungsw eis e=========================== @ 25 \mod_1404983235994_68.doc x @ 183019 @ 1 @ 1 →...
Page 15
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation Control switch Fig. 3: Right side of the freeze-dryer 10 Aeration valve 11 Access flange DN40 (e.g. for standard option "WTMplus”, behind the side panel) Fig. 4: Left side of the freeze-dryer Version 07/2022, Rev.
Page 16
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation 12 Name plate 13 Heat exchanger for the heat transfer medium (behind the panel) 14 Vacuum connection Fig. 5: Rear view of the freeze-dryer 15 Connection for a...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation 19 Vacuum sensor connection 20 Internal fuse 21 Power supply of the pressure control valve 22 Power supply of the vacuum pump 23 Automatic circuit breaker 1F1...
Page 19
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation Fig. 9: Vapour pressure curve for ice and water From a physical point of view, the freeze-drying process covers three phases (see figure below): (1) Freezing: The product to be dried is frozen under atmospheric pressure.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation Fig. 10: Freeze-drying phases Please find further information about basic principles, optimum procedures and applications in the brochure "Smart freeze-drying", which can be downloaded at www.martinchrist.de → [Applications] →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 2 Layout and mode of operation 2.2.2.4 Final drying Final drying is an option whenever one requires a product with minimal residual moisture. In the physical sense, this process is a desorption process, i.e.
Pos : 51 /200 C hrist/360 GT-BA Labor-Pilot (ST ANDARDM ODU LE)/030 Sic her heit/030-0010 Besc hilderung des Geräts GT @ 25\m od_1404983297040_68.docx @ 183215 @ 2 @ 1 → Marking of the unit The following symbols are used for Christ freeze-dryers: Dangerous voltage On (Power)
• ensure that the freeze dryer is operated with one of the vacuum pumps approved by Martin Christ Gefriertrocknungsanlagen GmbH (see chapter 5.2.3 - "Vacuum pumps"). • have the system maintained at regular intervals (see chapter 8 - "Maintenance and service").
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 3 Safety Qualified electrician Due to their special training, knowledge, experience and familiarity with the relevant standards and regulations, qualified electricians are in the position to perform work on electrical systems and to autonomously identify and prevent possible hazards.
(e.g. a special vacuum pump) and following a case-by-case examination and the written approval by Martin Christ Gefriertrocknungsanlagen GmbH! Risk of poisoning caused by azides in the products When loading and unloading the shelves, the personnel are exposed to the product.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 3 Safety Pos : 107 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/030 Sicherheit/030- 0020 Si cherheit Epsilon/030-0070- 0160 Gefahr en durch Überdr uck ohne Ü berdrucksic her ung ( Standar d) @ 36\m od_1435915684375_68.docx @ 266991 @ 3 @ 1 3.7.16 Hazards caused by overpressure in the chamber...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 3 Safety Pos : 111 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/030 Sicherheit/030- 0020 Si cherheit Epsilon/030-0091 Ver hal ten bei Gefahr en und Unfäll en Epsilon 1-4/2- 4 ++++++++++++++ @ 36\mod_1436336802184_68.docx @ 267237 @ 2 @ 1 →...
• Stop the freeze-dryer immediately in the event of a malfunction. Eliminate the malfunction (see chapter 7 - "Malfunctions and error correction") or contact the after-sales service of Firma Martin Christ Gefriertrocknungsanlagen GmbH (see chapter 7.3 - "Service contact"). •...
Page 39
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 3 Safety Fire prevention • Fuses protect certain electrical circuits within the freeze-dryer against over-current conditions. Always use fuses of the same type and rating. Safety area • Maintain a safety distance of at least 30 cm (12 inches) around the freeze-dryer.
→ 3.13 Remaining hazards All Christ freeze-dryers were built state-of-the-art and according to the accepted safety rules. Danger to life and limb of the operator, or of third parties, or impairments of the units or other material assets, however, cannot be completely excluded when the units are being used.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 4 Storage and transport Pos : 122 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/040 Lag erung und Trans port/040 Lager ung und Tr ansport=========================== @ 25\m od_1404983713664_68.docx @ 183467 @ 1 @ 1 →...
• Take the centre of gravity of the load in consideration (off-centre)! DANGER The freeze-dryer Epsilon 1-4 LSCplus weighs approx. 110 kg! The freeze-dryer Epsilon 2-4 LSCplus weighs approx. 140 kg ! CAUTION The freeze-dryer is packaged in a wooden crate.
The freeze-dryer Epsilon 1-4 LSCplus weighs approx. 110 kg! The freeze-dryer Epsilon 2-4 LSCplus weighs approx. 140 kg! CAUTION Freeze-dryers of the type Epsilon 1-4 LSCplus and Epsilon 2-4 LSCplus that are filled with natural, flammable refrigerants are not approved as air freight! NOTE •...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 4 Storage and transport Pos : 131 /200 Chris t/361 GT-BA Labor-Pil ot ( PROJ EKTE)/Epsilon 4er/Epsilon 1-4_2- 4 LSC plus Standar d/040 Lag erung und Trans port/040-0040- 0010 Tr ans portsic her ung en E1-4_2- 4 @ 25\m od_1404984722823_68.docx @ 186029 @ 3 @ 1 →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 5 Set-up and connection Pos : 134 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/050 Aufstellung und Ansc hluss/050 Aufstellung und Ans chl uss Epsilon/050 Aufstellung und Ansc hlus s Epsil on ========================== @ 36\mod_1436508805195_68.docx @ 267272 @ 1 @ 1 →...
The operating voltage on the name plate must correspond to the local supply voltage DANGER Christ freeze-dryers are units of safety class I. Epsilon 1-4 LSCplus and Epsilon 2-4 LSCplus units have a three-wire power cord with a CEE 7/4 plug (type F) (see chapter 10 - "Technical data").
Rotary vane pump Pascal 2010 C1 (only for Epsilon 1-4 LSCplus) Dry vacuum pumps: • Scroll pump Edwards nXDS 10iC (in the case of Epsilon 1-4 LSCplus only with an external power supply) • Scroll pump Pfeiffer HiScroll 12 Atex (only for Epsilon 1-4 LSCplus) •...
Page 48
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 5 Set-up and connection The vacuum pump is supplied with power by the unit, but the maximum current for the vacuum pump is limited. It is absolutely essential to refer to the label of the electrical outlet for the vacuum pump (see the following...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 5 Set-up and connection The freeze-dryer is equipped with a dry-running vacuum pump. As a result, the vacuum exhaust gases are completely free from oil particles. However, small quantities of debris from worn seals may be emitted. That is why,...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 5 Set-up and connection Pos : 154 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/050 Aufstellung und Ansc hluss/050 Aufstellung und Ans chl uss Epsilon/050-0020- 0070 Vak uumm esss onde neu - Ü berschrift--------------------------------- @ 48\mod_1490093443168_68.docx @ 492582 @ 3 @ 1 5.2.5...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Pos : 169 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/060 Betrieb/060-0050 Bel adetür Epsil on - Übersc hrift-------------------------------------- @ 25\m od_1404984628128_68.docx @ 185721 @ 2 @ 1 →...
Page 55
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Radiation Shield film for loading doors made of acrylic glass Loading doors made of acrylic glass have a special, transparent film on the exterior side of the acrylic glass door. This film reduces the transfer of radiant heat onto the product and ensures homogeneous freeze-drying results.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation The user interface is divided into four main windows that can be called up by touching the corresponding buttons: Manual This window is also the standard user interface. It is used to control the freeze-drying process manually.
Page 58
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Status line (1) This line shows the operating status of the freeze-dryer as well as the active phase. The status line also shows the current date and time. The clock is battery- buffered and must be reset after a failure (see chapter 6.6.1.3 - "Main...
Page 59
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Fig. 21: Dialog box "Select actual value" • Select the desired configuration or quit the dialog box by pressing the "return" button. In this way, it is possible to configure a personalised overview of the actual values.
Page 60
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Option: USB process recording (see chapter 6.7 - "Optional extensions and special equipment") • Select the "Process recording" function in the dialog box "Tools". • Select the input fields ("Batch data"). A keyboard for the data input will be displayed.
Page 61
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Fig. 24: Dialog box "Pressure increase test" • The "Test stop" button stops the pressure increase test. Option: LyoBalance weighing system If the freeze-dryer is equipped with a LyoBalance weighing system (see chapter 6.7 - "Optional extensions and special equipment"), then the...
Page 62
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Dialog box "Process and equipment messages" (7) This dialog box shows and saves all of the error messages and other messages. In the event of an error or message, the window “Process and equipment messages”...
Page 63
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Details Touching the message calls up details concerning the error message: • Cause of the message, • Effects of the message, • Measures to eliminate the error, → Reference designator, •...
Page 64
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Dialog box "Operating mode: select/start" (8) After the set values have been entered for the process, the process can be started manually with this function (see chapter 6.6.2 - "Manual freeze- drying").
Page 65
Fig. 31: Results of the performance test Evaluation: Please contact the Christ support team for the evaluation of the test run. Leakage test (2) The leakage test enables the chamber of the freeze-dryer to be tested for tightness in view of any gaseous or liquid media.
Page 66
• Select "Leakage test" and start the test via the "Run" button. The parameters for the leakage test have been developed by Martin Christ Gefriertrocknungsanlagen specifically for freeze-dryers. In a first step, these parameters (vacuum, ice condenser temperature) must be reached.
Page 67
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation At the end of the process, a pop-up window with a safety enquiry (confirmation prompt) opens. Following the confirmation of this enquiry, the freeze-dryer will be switched to standby. Evaluation: The test is passed if the dishes do not contain any water/ice after the end of the process.
Page 68
(confirmation prompt) opens. Following the confirmation of this enquiry, the freeze-dryer will be switched to standby. Evaluation: Please contact the Christ support team for the evaluation of the test run. Shelf mapping test (6) The shelf mapping test is a freeze-drying program for bringing the shelves to a temperature of +20°C, -40°C, and +40°C for a specific period of time in...
Page 69
The external measurement recording process can then be stopped. Evaluation: Please contact the Christ support team for the evaluation of the test run. Pos : 179 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1 Pos : 180 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/060 Betrieb/LSC plus /060- 0050-0010- 0010c Hauptfenster Manuell Labor/Epsilon T eil 3 @ 72\m od_1593154320311_68.docx @ 766569 @ @ 1...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Pos : 184 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/060 Betrieb/LSC plus /060- 0050-0010- 0031 H auptfens ter Optionen Epsilon - 1 Vakuumpum pe @ 72\m od_1593154328879_68.docx @ 766737 @ 4 @ 1 6.6.1.3...
Page 72
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Change scale unit (2) This dialog box is used to change the unit of measurement for the temperature and vacuum. Fig. 35: Dialog box "Change scale unit" Special equipment: Preselection of the aeration medium (3) This function can be performed in the manual mode or saved together with a program (see chapter 6.6.3.1 - "Creating a program").
Page 73
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation The following media preselection settings are possible: • Air: Preselects the media connection "Air". • Inert gas: Preselects the media connection "Inert Gas". • Automatic: Automatic media connection preselection depending on the chamber pressure and stoppering status of the product vials: –...
Page 74
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Special equipment: Inertisation (5) (see chapter 6.7 - "Optional extensions and special equipment") If the "Inertisation" option is active and the freeze-dryer is in the "Standby", "Manual - Freezing", "Program - Loading" or "Program - Freezing" mode, the automatic inertisation of the drying chamber will be performed for the preselected inertisation time after the loading door has been closed.
Page 75
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation If the volume flow through the inertisation line falls below a certain threshold for more than 5 seconds (the value can be preset at the flow meter) or if the loading door is opened during an inertisation sequence, the process will also be aborted immediately.
Page 76
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Network (2) This dialog box is used to change various system settings. The modifications will not become effective until after a restart of the unit. NOTE Fig. 41: Dialog box "Network"...
Page 77
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Click on touch: If this function is active, a clicking sound can be heard whenever the system registers that a button has been touched. Behaviour of the pressure control valve in the case of insufficient ice condenser cooling: If this function is active (button "closed"), the pressure...
Page 78
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation • The button with the lock symbol blocks the access immediately and the system switches to the default setting. Fig. 45: Access blocked, the data cannot be edited Operator/maintenance/administrator password: For each of these access levels, certain editing rights have been defined.
Page 79
• Press the button "Add option". An input window opens. • Enter the six-digit Christ activation code that was supplied for this option. Note that the keys are case sensitive. Options can be removed in the same way. The modifications will not become effective until after a restart of the unit.
Page 80
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Product temperature sensor (see the illustration above, item B) Only for systems with a WTMplus system The wired product temperature sensor 1 is the default sensor for determining the value "ΔT product" (see chapter 6.6.3 - " Freeze-drying with the PGMplus programmer module ", section "Conditions for switching...
Page 81
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation WTMplus configuration (7) The dialog box is displayed for information purposes only. The configuration can be changed only via the LPCplus user interface. NOTE This dialog box shows all of the existing WTMplus sensors. Every sensor is assigned to a measuring channel based on its serial number.
Page 82
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Maintenance "Service" options: "Operating hours" "Failure memory" "Vacuum pump oil change" "Maintenance" "Sensor adjustment" Fig. 50: Dialog box "Service" (varies depending on the type of system) Operating hours (1) This dialog box is used to call up the number of operating hours of the various components of the freeze-dryer, e.g.
Page 83
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Failure memory (2) Equipment information system. These messages can be viewed in this dialog box. The failure memory includes the last 32 messages. If this number is exceeded, the oldest message will be overwritten.
Page 84
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Vacuum pump oil change (3) This option only applies to oil-sealed vacuum pumps! NOTE The system monitors the oil change interval of the vacuum pump. The interval can be adapted to the vacuum pump model and utilisation.
Page 85
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Maintenance (4) The maintenance interval of the unit is fixed at 3,000 operating hours or at least one maintenance per year. When the end of a maintenance interval is reached, a corresponding message will be displayed.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation After the completion of a phase, the system switches to the next phase without switching the unit to standby. The transition from "freezing" to "warm-up" takes place automatically. After the completion of the warm-up phase, this dialog box will be displayed: Fig.
Page 88
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Viewing or editing the set values: • Press the button "Set values: view/edit" (see chapter 6.6.1.1 - "Main window "Manual""). The following dialog box will be displayed: "Aeration" button "Defrosting" button "Media drain"...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Numerical values can be edited with the aid of a numerical keypad: Set value display Button for changing the sign (e.g. when entering temperature values) "" (infinite) Button for editing the...
Page 90
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Fig. 59: Representation of a freeze-drying program Program sections When the programmer module executes a freeze-drying process, it executes the various sections that were created one after the other until the last section is completed.
Page 91
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Conditions for switching to the next section For all of the program sections to be executed automatically, certain switching conditions must be fulfilled at the end of each section. If these conditions are not fulfilled, e.g.
Page 92
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation T comparative (option): This value indicates the difference between a Pirani sensor and a capacitive vacuum sensor in per cent based on the actual value of the capacitive vacuum sensor. When the sublimation rate decreases, the difference decreases as well.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Pos : 194 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/060 Betrieb/LSC plus /060- 0050-0032- 0011 Programm erstellen Epsilon @ 72\m od_1593154338701_68.docx @ 766927 @ 4 @ 1 6.6.3.1...
Page 94
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation "Program" buttons: "Program" "Program name" "Show diagram" "Insert section" "Delete section" Button for calling up more functions and set values "Save" Fig. 63: Editing a blank program template • The button “Delete section” (5) can be used to delete sections.
Page 95
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Button ">>more" The button ">>more" can be used to enter more functions and set values in a program-related manner. Fig. 64: Selection of program-related functions and set values Vacuum pump warm-up see chapter 6.6.3 - "...
Page 96
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation next section is fulfilled and the next section will be initiated. If the condition for switching to the next section is not fulfilled, a process message will be issued, the current section will be extended, and further pressure increase tests will be executed periodically until the condition is fulfilled.
Page 97
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation → Option: Comparative pressure measurement (see chapter 6.7 - "Optional extensions and special equipment") The indication of the measurement value "dp Comp" can be configured in chapter 6.6.1.1 - "Main window "Manual"" under "Actual values of the current process"...
Page 98
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Option: LyoBalance weighing system If the freeze-dryer is equipped with a LyoBalance weighing system (see chapter 6.7 - "Optional extensions and special equipment"), then the corresponding parameters must be configured in this dialog box.
Page 99
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation "Aeration, (stoppering), storage" procedure: 1. Aeration up to the desired partial vacuum ("Aeration" value): After the actual freeze-drying process has been completed, the ice condenser continues to be cooled and the shelves are held at the temperature of the last drying section.
Page 100
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Special equipment: Preselection of the aeration medium If the freeze-dryer is equipped with an automatic aeration valve and automatic valves for media selection, this dialog box can be used to preselect the medium that is to be used for the aeration medium requests during the execution of the particular program.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Program templates (2) The PGMplus programmer module offers various program templates that include recipes for all kinds of freeze-drying applications. They are suggestions for orientation purposes and must be adapted to the specific area of application of the freeze-dryer.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Fig. 71: The freeze-drying process can be started with the aid of the button “Program start”. • During the freeze-drying process, the description of the active program can also be called up from the main window "Program" and via the button "Info".
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Pos : 204 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/060 Betrieb/LSC plus /060- 0062 Opti onale Erw eiterungen und Sonder auss tattung en Epsilon Allgemei n ----------------- @ 72\m od_1593154343014_68.docx @ 767041 @ 2 @ 1 Optional extensions and special equipment The freeze-dryer can be equipped with the following optional equipment.
Page 105
Option: LPCplus SCADA System The Christ LPCplus system consists of the Supervisory Control And Data Acquisition (SCADA) software program by Martin Christ Gefriertrocknungsanlagen GmbH and a dedicated PC. The system is...
Page 106
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Option: USB process recording see also chapter 6.6.1.1 - "Main window "Manual"", section "Dialog box "Tools"" This feature enables the recording of a running process on a USB storage medium. After the end of the process recording, the process data can be viewed on the PC with LyoLogplus and they can be printed.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 6 Operation Switching the freeze-dryer OFF The freeze-dryer must be in the standby status. • Switch the freeze-dryer off by pressing the mains switch. Pos : 207 /010 Univ ers almodul e/ Abs chnittswec hs el @ 0 \mod_1202124514062_0.docx @ 418 @ @ 1...
(see chapter 7.3 - "Service contact") If it is impossible to eliminate the errors, contact the Christ service department! NOTE Pos : 212 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1 Pos : 213 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/070 Störungen und F ehl ersuche/070- 0010- 0010 Netzaus fall @ 25\mod_1404984058964_68.doc x @ 184153 @ 3 @ 1...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 234 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080 Wartung und Instandhaltung Al pha,Beta,Gamm a,Epsil on 1- 4/2-4 =========================== @ 60\m od_1540462045071_68.docx @ 602079 @ 1 @ 1 →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Cleaning of the freeze-dryer Risk of burns on hot surfaces Housing of the freeze-dryer During the operation of the freeze-dryer and half an hour afterwards, the outer surface of the freeze-dryer (especially the pipes and units) may be WARNING hot.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 241 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0010- 0022-0010 F olienbesc hichtung der Beladetür Epsilon Standar d 2.0 @ 92\m od_1658308697576_68.docx @ 967406 @ 4 @ 1 8.1.2.1...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 243 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0010- 0022-0020 T üversc hl ussriegel Epsilon @ 25\mod_1404984677136_68.doc x @ 185889 @ 4 @ 1 →...
Check the heat exchanger at least once per month for soiling and clean it if necessary. • Please contact the Christ service department if you have any queries (see chapter 7.3 - "Service contact"). Pos : 248 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1...
The number of operating hours of the vacuum pump can be monitored by way of the LSCplus control system and also by way of the CHRIST SCADA software LPCplus (see chapter 6.6.1.3 - "Main window "Options""/"Operating hours"...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 257 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0010- 0120 Vak uumm esss onde allgemei n @ 25\m od_1404984105610_68.docx @ 184433 @ 3 @ 1 →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 261 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0020 D esi nfektion von Tr ock nungskamm er und Z ubehör @ 25\m od_1404984107851_68.docx @ 184461 @ 2 @ 1 →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 263 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0030 W artungs plan Epsil on 4er (Vak uum pumpen mit Öl) @ 25\m od_1404984679421_68.docx @ 185916 @ 2 @ 1...
Page 126
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service System/ Detail / Part Inspection Periodical Maintenance Repair / Rebuild Component Cooling water Cleaning of the valve controller Heat exchanger Cleaning Mechanical Vial stoppering components: system operation Leakages...
Page 127
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service System/ Detail / Part Inspection Periodical Maintenance Repair / Rebuild Component Switch cabinet: Replacement of the 1 year filter cooling system Wiring Power readings of components LSCplus control...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 265 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0031 W artungs plan Epsil on 4er (Troc kenläuferpum pe) @ 85\mod_1631771127780_68.doc x @ 897660 @ 2 @ 1...
Page 130
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service System/ Detail / Part Inspection Periodical Maintenance Repair / Rebuild Component Ice condenser: temperature low end Cooling water Cleaning of the valve controller Heat exchanger Cleaning Mechanical Vial stoppering...
Page 131
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service System/ Detail / Part Inspection Periodical Maintenance Repair / Rebuild Component Switch cabinet: Replacement of the 1 year filter cooling system Wiring Power readings of components LSCplus control...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service Pos : 267 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/080 W artung und Instandhaltung/080-0041 Ins tandhaltungs arbeiten Epsilon (mit W artungsv ertrag nur de) @ 25\m od_1404984109982_68.docx @ 184489 @ 2 @ 1...
Page 133
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 8 Maintenance and service 1. Declaration of decontamination As a certified company and due to the legal regulations for the protection of our employees and of the environment, we are obliged to certify the harmlessness of all incoming goods.
Pos : 273 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMODULE)/090 Entsorgung/090-0010 Entsorgung der Gefriertr ock nungsanlag e @ 25\m od_1404984141122_68.docx @ 184531 @ 2 @ 1 → Disposal of the freeze-dryer Martin Christ Gefriertrocknungsanlagen GmbH is a registered manufacturer of electric and electronic devices that are solely intended for commercial use.
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 10 Technical data Pos : 278 /200 Chris t/361 GT-BA Labor-Pil ot ( PROJ EKTE)/Epsilon 4er/Epsilon 1-4_2- 4 LSC plus Standar d/100 T ec hnis che D aten/100 Tec hnisc he D aten E1- 4_2-4 Standard 2.0 =================================== @ 25 \mod_1404984770073_68.doc x @ 186141 @ 1 @ 1 →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 11 Appendix 11.2 EC declaration of conformity in accordance with the EC Machinery Directive Version 07/2022, Rev. 3.0 of 20/07/2022 • sb 139 / 152 Translation of the original operating manual...
Page 140
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus 11 Appendix Pos : 290 /010 Univ ers almodul e/ Seitenwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1 Version 07/2022, Rev. 3.0 of 20/07/2022 • sb 140 / 152 Translation of the original operating manual...
Since absolutely tight components simply do not exist, a leak rate is determined. The parameters for the leakage test have been developed by Martin Christ Gefrier- trocknungsanlagen GmbH specifically for freeze-dryers. In a first step, these parameters (vacuum, ice condenser temperature) must be reached.
Page 146
Safety pressure Since the vacuum has a dominating influence on the product temperature, Martin Christ Gefriertrocknungsanlagen GmbH has integrated a so-called safety pressure feature into the freeze-dryers in order to ensure the protection of the product. If the pressure inside the drying chamber increases too strongly so that it exceeds the safety limit, the energy supply of the shelves will be interrupted and the →...
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Index 13 Index CE mark in compliance with the directive 2006/42/EC ..........23 Access protection........77 Chamber, cleaning and care ....115 Accessories ....... 39, 114, 122 Changing of operating supplies ....38 Accident prevention ......
Page 148
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Index Danger of infection due to exhaust air ..49 Editing a program........101 Danger of poisoning due to exhaust air ..49 Electrical connection ........ 135 Dangerous materials ........ 123 Electrical system (maintenance) ....118 Data interface (LAN) ........
Page 149
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Index Gas injection with a proportional valve 73, 106 Ice condenser .......... 135 General conditions ........12 Ice condenser temperature ......91 General information on freeze-drying ..18 Importance of the operating manual ..... 9 General malfunctions .......
Page 150
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Index Main window "?" ......... 86 Main window "Manual" ....... 57 Oil mist separator ........48 Main window "Options" ......71 Oil mist separator (exhaust filter) and vacuum Main window "Program" ......70 pump (oil-sealed) ........
Page 151
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Index Power failure ..........109 Safety area ..........39 Power supply ..........46 Safety data sheets ........136 Preparation .......... 20, 53 Safety devices ........... 36 Preselection of the aeration medium .. 72, 106 Safety distance ..........
Page 152
Freeze-dryer Epsilon 1-4 LSCplus Freeze-dryer Epsilon 2-4 LSCplus Index Storage ............41 UV radiation ........45, 114 Storage and transport ........ 41 Storage conditions ........41 Vacuum ............. 91 Stress-corrosion ........122 Vacuum connection ......... 136 Sublimation ..........146 Vacuum pump (dry-running) and silencer .. 49 Supply voltage ...........