Clean the surface with a cloth and ethanol
Clean exterior
(sides, back,
>70% (v/v).
front)
Task
Clean
touchscreen
Clean USB slots
Clean ethernet
slot
Clean Pneumatic
Printhead
Clean
Temperature-
controlled
Printhead
BIO X6 User Manual v. 1.0
Action
Clean carefully with a smooth cloth,
avoid using harsh solvents.
No chemical solution or other liquid
should be used when cleaning. Use a
dust blower to carefully remove dirt and
dust.
No chemical solution or other liquid
should be used when cleaning. Use a
dust blower to carefully remove dirt and
dust.
Clean the printhead with a smooth cloth
to remove any dust and debris. When not
being used, store the printhead inside a
sealed plastic bag in a cool and dry place
to avoid dirt accumulation or
deterioration of the instrument.
Clean the printhead body with a smooth
cloth to remove any dust and debris.
Check the fan of the printhead for
residues or accumulated dirt. Remove
the printhead from the BIO X6 and use a
dust blower
to carefully eliminate unwanted dust
particles accumulated on the fan. When
not being used, store the printhead
inside a sealed plastic bag in a cool and
dry place to
Every week
Regularity
Check
Dates
Every week
Every 3
months
Every 3
months
Every time
it is used
Every time
it is used
144