MET1000-DE All in One True RMS Electrical Tester User Guide...
Page 2
Megger Limited, Archcliffe Road, Dover, Kent CT17 9EN. ENGLAND T +44 (0)1304 502101 F +44 (0)1304 207342 www.megger.com Megger Ltd reserves the right to alter the specification of its products from time to time without notice. Although every effort is made to ensure the accuracy of the information contained within this document it is not warranted or represented by Megger Ltd.
2014/53/EU. Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014/30/EU and 2014/35/EU where they apply. The full text of Megger Instruments EU declarations of conformity are available at the following internet address megger.com/company/about-us/eu-dofc www.megger.com...
Category III or IV according to IEC 61010-031 and shall have a voltage rating of at least the voltage of the circuit to be measured. Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity. MET1000-DE www.megger.com...
Page 5
When connecting the test leads to the device under test connect the common test leads before connecting the live test lead. When removing the test leads, remove the live test lead before removing the common test lead. www.megger.com MET1000-DE...
Page 6
Timing Rating (tr): 30 seconds, Recovery Time (rt): 240 seconds, when measuring >300 V, recovery time is necessary. L/R LED may light up when measuring AC voltage. Due to the high internal resistance, capacitive and inductive voltage (ghost voltage) may be indicated. MET1000-DE www.megger.com...
Page 7
If a voltage appears on the screen then use 2 Pole method to prove the circuit live. www.megger.com MET1000-DE...
Counter-clockwise Phase Sequence L1-L3-L2 (Left) Source Terminal Grip firmly Warning Phase Rotation Test works only on 3 phase 4 wire systems. The result is unreliable on any other systems. Note It maybe necessary to check the result by testing in reverse sequence. MET1000-DE www.megger.com...
Do not hold the meter above the barrier Warning The barrier guard on the meter provides maximum protection when holding the unit in normal use. Warning Remove the test leads from the back of the meter when measuring current. MET1000-DE www.megger.com...
Using the Functions Using the Functions HOLD Press Manually powering down the meter Press Ω/A button and “HOLD” button at the same time. Powering ON www.megger.com MET1000-DE...
Page 12
Using the Functions Torch Press >2 sec Powering ON Self-diagnostic test Press Warning Do not use the tester if a fault is indicated during self-diagnostic test. MET1000-DE www.megger.com...
Page 13
Using the Functions Function Button Voltage/Continuity/Single Pole Mode Press Button Press Button >2 sec Resistor Mode Press Button Current Mode www.megger.com MET1000-DE...
Battery Replacement Battery Replacement Press here Press here Warning Replace the batteries as soon as the low battery indicator starts to flash to avoid false readings. Remove test leads from meter before opening the battery cover. MET1000-DE www.megger.com...
Without the probe tip guard cap the test leads can ONLY be used in a CAT II environment. Make sure the test leads are firmly inserted in the instrument and other accessories are connected securely. www.megger.com MET1000-DE...
Test lead assembly Test lead assembly Warning Do not attach test leads onto the back of the meter while measuring current. Press here For 19 mm power socket For most applications except current measurement. MET1000-DE www.megger.com...
-20 to +60 °C, 0 to 80 % RH. (no batteries) Temperature coefficient : 0,2 x (specified accuracy) / °C, < 18 °C, > 28 °C IP rating : IP65 Vibration : Random Vibration per MIL-PRF-28800F Class 2 Drop Protection : 1.2 m drop to hardwood on concrete floor www.megger.com MET1000-DE...
Page 18
Accuracy by Crest Factor (C.F.): Add 1.0% for C.F. 1.0 ~ 2.0 Add 2.5% for C.F. 2.0 ~ 2.5 Add 4.0% for C.F. 2.5 ~ 3.0 Max. Crest Factor of Input Signal: 3.0 @ 5000 counts 1.5 @ 10000 counts MET1000-DE www.megger.com...
Page 19
45 Hz to 400 Hz Max. Input Current : < 3.5 mA @ 1 000 V Overload Protection : AC/DC 1,000 V Resistor Range Resolution Accuracy 9999 Ω 1 Ω ±(1.5 % + 5D) 50.00 kΩ 0.01 kΩ Output Voltage : About 0.5 V Overload Protection : AC/DC 1,000 V www.megger.com MET1000-DE...
Page 20
“L” LED is on when the signal of L2 probe lead the signal of L1 probe; “R” LED is on when the signal of L1 probe lead the signal L2 probe. Single-Pole Phase Check Sensitivity : 90 V to 1000 V (Phase-to-Earth) Frequency Range : 45 Hz to 65 Hz Indicator : ELV LED MET1000-DE www.megger.com...
The crossed out wheeled bin symbol on the instrument and on the batteries is a reminder not to dispose of them with general waste at the end of their life. Megger is registered in the UK as a Producer of Electrical and Electronic equipment. The registration No is;...
Page 22
Some states or countries laws vary, so the above limitations or exclusions may not apply to you. Megger Limited Archcliffe Road Dover...