EXPLORER BALANCES
5.6

Balance Setup

Enter this sub-menu to customize Balance display functionality.
5.6.1

Balance Set-up sub-menu

Stability
Indicator Range
Gross Indicator
Factory default settings are shown below in bold.

Filter Level

Legal For Trade
(for Approved
models)
5.6.2

Stable Range

Set the amount the reading can vary while the stability symbol remains on.
0.5 D
1 D
2 D
3 D
5 D
Note: When LEGAL FOR TRADE (below) is set to ON, the setting is forced to 1 D.
The setting is locked when the Security Switch is set to the ON position.
5.6.3
Filter level
Set the amount of signal filtering.
LOW
MEDIUM
HIGH
The setting is locked at the current setting when the security switch is set to
the ON position.
5.6.4

Auto Zero Tracking

Set the automatic zero tracking functionality.
OFF
0.5 D
1 D
3 D
Note: When Legal for Trade is set to ON, the AZT setting is forced to 0.5 D.
The OFF setting is still available. The setting is locked at the current setting when the
security switch is set to the ON position.
5.6.5

Audible Signal

Set to turn on a beep sound when weighing is stable.
OFF = disabled.
Stable Weight Only = enabled the beep sound when weight is stable
Auto Zero
Audible Signal
Tracking
Graduations
= 0.5 graduations
= 1 graduation
= 2 graduations
= 3 graduations
= 5 graduations
= faster stabilization time with less stability.
= normal stabilization time with normal stability.
= slower stabilization time with more stability.
= disabled.
= display maintains zero up to a drift of 0.5 graduation
per second
= display maintains zero up to a drift of 1 graduation per
second.
= display maintains zero up to a drift of 3 graduations per
second.